HP 8901B Operation And Calibration Manual page 333

Hide thumbs Also See for 8901B:
Table of Contents

Advertisement

Model
8901B
Operation
23.
External
LO
(Option
030
Only).
Use this special function when making Selective Power
Measurements using an external LO. (Refer to Selectiue
Power
Measurements on page
3-95
for more
information about this measurement.)
23.0
Return the
LO to
internal.
23.1
Set the LO
to
external.
23.2
Display the status of the
LO (O=Internal, l=External).
24.
Selective Power Measurement (Option
030
Only).
Use this special function when measuring
adjacent channel power or single-sideband noise. (Refer to Selective Power Measurements on page
3-95
for more information about this measurement.)
24.0
24.1
24.2
24.3
24.4
24.5
24.6
24.7
24.8
24.9
Sets the Modulation Analyzer
to
the Selective Power Measurement Mode.
Selects the wide-bandwidth
IF
filter for adjacent-channel or cellular radio power measure-
ments; displays the
IF
level (in volts).
Establishes the level measured with
24.1
SPCL as the 0
dB
reference.
Selects a narrower bandwidth
IF
filter than
24.1
SPCL
for adjacent-channel power
measurements; displays the
IF
level (in volts).
Establishes the level measured with
24.3
SPCL as the 0
dB
reference.
Selects a very narrow bandwidth
IF filter
(2.5
kHz)
for the single-sideband noise measure-
ment; displays the
IF
level (in volts).
Establishes the level measured with
24.5
SPCL
as
the 0
dB
reference.
Displays the noise measurement as normalized
to a
1
Hz bandwidth.
Enables setting of the value of the noise measurement bandwidth.
Displays the value of the noise measurement bandwidth.
25.
Current Value
of
External Attenuation.
Use
this special function in conjunction with
dB EXT
ATTEN (Blue Key, LOG/LIN)
to
scale level measurements
to
correctly indicate the power
at
the
input of an external attenuator or amplifier connected
to
the Modulation Analyzer. (For information
on setting the value of external attenuation, refer to Erternal Attenuation on page
3-105.)
25.2
Display the current offset in
dB.
(Amplification is shown as a negative attenuation value.)
27.
Frequency Offset Control.
Use this special function when
making
measurements on down-
converted input signals. (The Modulation Analyzer can measure frequencies to
42
GHz when an
external mixer and
LO
are used
to
down-convert the input signal. Refer
to
l+equency offset Control
on page
3-225
for requirements of the local oscillator and mixer.) The Requency Offset Control
special function allows the Modulation Analyzer to keep track of the frequencies involved in the
down-conversion.
27.0
Exit the Frequency Offset Mode.
27.1
Enter the Frequency Offset Mode.
27.2
Display the external
LO
frequency.
27.3
Prepare to enter and enable the
LO
frequency. After entering
27.3
SPCL, enter the frequency
of the LO (in MHz). Flrequencies from 0 to
40700
MHz can be entered.
29.0
SINAD Measurement.
Use this special function to make
SINAD
(Signal to Noise And
Distortion) measurements on either internally demodulated signals or externally applied audio signals.
r e a l
5SEP88
3-245