Determine The Problem - KERN MPE 250K100HNM Service Manual

Personal floor scale with bmi function
Hide thumbs Also See for MPE 250K100HNM:
Table of Contents

Advertisement

Err 12
Err 13
Err 14
Err 15
Err 16
Err 17
Err 18
Err19
Err P
Err l
--ol--
--lo--
Fai l h /
fai l l / fai
l
Ba lo /
lo ba

8.3 Determine the Problem

Determine whether the problem is in the PCB or the Load Cell
 Disconnect the power supply from the scale, and disconnect the load cell
connection from the PCB
 Reapply power and test the PCB
 If problem goes away, its source is probably in the Load cell. Check the wiring,
connecter, load cell and mechanical components of the load cell.
 If problem persists, its source is probably in the PCB. Check the PCB
voltages, connecters, cables and function programs
20
Accumulation error
Lack of unit weight
Lack of sample
Gravity error
Paper error
Tare out of range
Pre-tare error
Initialize
zero error
Printer error
Approval setting error
Over range
Underload
Calibration Error
Battery low
Max accumulation times 99 /
999 / weight 999999
Check unit weight entry data,
must be >0.5d
Check counting samples
entry, must be >20d
Check the gravity settings.
Gravity range must be 0.9xx ~
1.0xx
Check the printer paper
Minus weight or overload.
Remove the load and restart
scale again.
Check the pre-tare value
Calibration the scale.
Check the printer settings or
connections
Check the PCB jumper
settings. Must be connect
jumper pin to K1 (BW series)
Remove the load.
Re calibrate
Minus weight, check the
platform and restart or
calibrate.
Check the test weights & Re
calibrate
Re charge battery, check the
voltages.
MPE_HNM / MPE_PNM -SH-e-1810