Overview:
Function
F0 di
Modification of readability
F1 az
Automatic zero point
correction
(zero tracking)
F2 bt
F3 sp
F4 ip
F5 ut
F6 sl
Standby mode
see chap. 4.3.1
F7 gv
F8 ci
Password entry
F9 CL
* = default setting
24
Available
Description
settings
Low
High*
d (low)
cap
1 t
2kg
1kg
3 t
10kg
5kg
5 t
10kg
5kg
10 t
20kg
10kg
AZn 0
0.5 d
AZn 1*
1 d
AZn 2
2 d
AZn 3
4 d
Not documented
Not documented
Internal A/D converter value
Not documented
SLP 0
Standby mode switched off
SLP 1*
Standby mode after 5 minutes
SLP 2
Standby mode after 10 minutes
SLP 3
Standby mode after 20 minutes
SLP 4
Standby mode after 30 minutes
Not documented
In „P1- - - " display increase the flashing number using
, select number using HOLD. Confirm entry by
ZERO button.
Adjustment, see chapter 7
d (high)
500g 500g 200g 100g
2kg
2kg
1kg
2kg
2kg
1kg
5kg
5kg
2kg
HFM-BA-e-1813
500g
500g
1kg