3. Procedure to adjust an isotope dial setting in the Atomlab 500 Dose Calibrator 4. Procedure to change one of the button isotopes to be Ra-223 086-330 086-332 IODEX Biodex Medical Systems, Inc. 20 Ramsey Road, Shirley, New Y ork, 11967-4704, Tel: 800-224-6339 (Int’l 631-924-9000), Fax: 631-924-9241, Email: info@biodex.com, www.biodex.com FN: 13-305 11/13...
Page 2
18. Select YES to save the new isotope NOTE: You have now added Ra-223 to your isotope list. 2. Procedure to set a dial value for Ra-223 for the Atomlab 500 If the user has a Ra-223 NIST traceable standard sent from the manufacturer, they can place the standard into the dose calibrator chamber and verify the factory determined isotope setting or create their own custom dial value.
Page 3
3. Procedure to adjust an isotope dial setting in the Atomlab 500 dose calibrator Press the HOME key Press SELECT UTILITIES Select DOSE CALIBRATOR CONFIGURATION Select ISOTOPE LIST – you must now enter the password The ISOTOPE LIST screen is now displayed Select CHANGE DIAL VALUES;...
Page 4
Certified Quality Management System IODEX M A D E I N Biodex Medical Systems, Inc. 20 Ramsey Road, Shirley, New Y ork, 11967-4704, Tel: 800-224-6339 (Int’l 631-924-9000), Fax: 631-924-9241, Email: info@biodex.com, www.biodex.com U . S . A .