Phcbi IncuCyte MCO230AICUV Instructions For Use
Hide thumbs Also See for IncuCyte MCO230AICUV:

Advertisement

Instructions For Use:
MCO230AICUV - IncuCyte kit

Installation

1. Follow the installation instruction from the original instructions for use of the MCO-230AIC-PA.
2. Take out the original humidifying pan, humidifying pan cover and the 4 trays (see Fig. 1 and 2).
3. Place the smaller humidifying pan
with part number 624 230 4548 on
the bottom of the incubator and
make sure to place it with the long
side against the back wall of the
incubator so water accumulated on
the dew stick will drip in this
humidifying pan.
4. Place the complete humidifying pan cover with part number 624
230 4555 over the humidifying tray and make sure to place it
against the back wall of the incubator.
5. Place the stand, NU-STAND-230AICGEN, for the IncuCyte S3 system on the bottom of the incubator
6. Install the IncuCyte S3 inside the incubator. Set temperature to 36.5 °C (see manual of MCO-230AIC-
PA to adjust).

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the IncuCyte MCO230AICUV and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Phcbi IncuCyte MCO230AICUV

  • Page 1 Instructions For Use: MCO230AICUV - IncuCyte kit Installation 1. Follow the installation instruction from the original instructions for use of the MCO-230AIC-PA. 2. Take out the original humidifying pan, humidifying pan cover and the 4 trays (see Fig. 1 and 2). 3.
  • Page 2 Adjustment heater balance 1. Go into the service mode of the incubator by pressing the Menu key more than 5 sec. at top display to move to Service code display. 2. Input “384” at Service code display and press OK key.
  • Page 3 4. Press Heater Balance key. Hum Heater: Bottom heater Door Heater: Door heater 5. Adjust the heater balance. Set the Hum Heater to 1 Set the Door Heater to 40 Press Apply key. 6. Press Top key to return to the top display.