Nonin Onyx II 9550 Instructions For Use Manual page 8

Finger pulse oximeter
Table of Contents

Advertisement

Immunity Test
Electrostatic Discharge (ESD)
IEC 61000-4-2
Power Frequency (50/60 Hz) Magnetic Field
IEC 61000-4-8
Radiated RF
IEC 61000-4-3
Harmonic Emissions (IEC 61000-3-2), Voltage Flicker Emissions (IEC 61000-3-3), Electrical Fast Transients (IEC 61000-4-4), Surge (IEC
61000-4-5), Voltage dips (IEC 61000-4-11), Conducted Immunity (IEC 61000-4-6)
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
Table 2: Electromagnetic Immunity
80 MHz – 2.7 GHz
Table 3: Not Applicable
Compliance
±8 kV contact
±15 kV air
30 A/m
380 – 390 MHz
430 – 470 MHz
704 – 787 MHz
800 – 960 MHz
1.7 – 1.99 GHz
2.4 – 2.57 GHz
5.1 – 5.8 GHz
10 V/m
27 V/m
28 V/m
9 V/m
28 V/m
28 V/m
28 V/m
9 V/m
8