IV. 6. Technical characteristics
Dimensions: D x W x H: 368 x 419 x 699 mm.
Weight: 26.4 kg.
Noise level: ±55 dBA
Flow values:
0 - 10 l/min.
Average oxygen content:
• at 2 l/min. 92%.
• at 9 l/min. 91%.
• at 10 l/min. 90%.
(values at 21 °C and at an atmospheric pressure of 1013 mbar). Max. recommended flow:
10 l/min. The maximum outlet pressure is 138 kPa. It is not recommended to use the NewLife
Intensity 10 at flow values less than 2 l/min. unless utilized with the Pediatric Option Procedure.
Electrical power supply:
• 220 - 240 V - 50 Hz Europe / 115 V – 60 Hz / 220 - 240 V – 60 Hz (other
countries, depending on version)
• Use the cable provided, and check that the electrical characteristics of the mains socket used
match those indicated on the manufacturer's plate on the rear panel of the machine.
• Mean power rating: 590 watts
Class II
Type BF
3.0 amps (220 - 240 V).
6.0 amps (120 V).
Filters:
At the rear of the device: a dust filter.
Before the oxygen outlet: a product filter < 2.0 µm.
Air circulation:
A fan cools the compressor compartment.
Environmental limit conditions:
The performances of the device (especially the oxygen concentration) are quoted at 21° C and
1013 mbar. They may change with temperature and altitude.
- The device should preferably be stored and transported in the vertical position.
- The device must only be used in the vertical position.
- Ambient temperature of between 5° C and 35º C (operation).
- Storage temperature range from -20° C to 60° C.
- Relative humidity of between 0% and 95% (operation and storage).
- IPX1: Protected against dripping water (complies with the EN 60601-1 standard)