DIN TU - in parking position
DIN TU - connected
7.3.2. DISCONNECTION OF TU - OTHER TYPES (CZ, BSI, SS)
To disconnect the probe (L2) from TU
L2
B
hold the probe and press the button
(B) into the final position. Probe (L2) is
spontaneously (by the gas pressure)
pulled out from TU.
During disconnecting hold the probe
firmly to prevent the probe springing
out of the TU. Note danger of injury.

7.4. AFTER EVERY USE

TU - connected
Disconnect all medical devices from the
TU after every use!
8. CLEANING
Remove dirt with a soft cloth damped in oil free oxygen compatible soap
water and rinsed with clean water. Disinfection can be carried out with an
alcohol-based solution (spray or wipes).
If other cleaning solutions are used, check that they are not abrasive and
that they are compatible with brass, plastic materials of components, and
gas.
Do not use cleaning solutions containing ammonia!
Do not immerse in water or any liquid.
Do not expose to high temperature.