7820A GC Site Preparation
18
Table 10 Delivery pressures required at the GC, in kPa (psig) (continued)
Detector type
FID
NPD
Carrier
(max)
Carrier (min)
Conversions: 1 psi = 6.8947 kPa = 0.068947 Bar = 0.068 ATM
Notes:
• If you have not requested option 305, you must supply
pre-cleaned, 1/8-inch copper tubing and a variety of 1/8-inch
Swagelok fittings to connect the GC to inlet and detector gas
supplies.
• If you purchased automated valving, the valve actuation
requires a separate pressurized, dry air supply at 380 kPa
(55 psig). This air supply must end in a male fitting
compatible with a 1/4-inch id plastic tube at the GC.
• Never use liquid thread sealer to connect fittings. Never use
chlorinated solvents to clean tubing or fittings.
Inlet type
TCD
uECD
Split/Splitless Purged
827 (120)
(20 psi) above pressure
used in method
7820A GC Site Preparation Guide
packed
827 (120)