E+E Elektronik EE1900 User Manual page 10

Humidity module for climate chambers
Hide thumbs Also See for EE1900:
Table of Contents

Advertisement

U [ V DC ]
5 - 30
T [°C (°F)]
160 (320)
ambient temperature
start automatic
recovery ARC
U [ V DC ]
5 - 30
Fig. 5
ARC Chart
During the sensor recovery process, the output is frozen at the last measured value before triggering
the sensor recovery.
The ARC mode is indicated by the closed contact at the optocoupler output (No. 3 „Status" - Fig. 3)
and by the red LED flashing very quickly. The optocoupler output thus indicates wether the data on
the analogue output is up-to-date (during normal measurement) or frozen (during the sensor recovery
process).
An efficient sensor recovery requires a certain sensor temperature. In case of strong air flow at the
sensing head (for instance more than 6 m/s with the stainless steel grid filter), the required sensor
temperature might not be reached. In such a case, the heating process is automatically interrupted
after 2 minutes.
10
Heat IN
min. 3 s
ARC T Profile
heating time 20 min
freezing period of output signals
ARC Status
User Manual EE1900 Humidity Module for Climate Chambers
cooling time min. 3 min
T [min]
T [min]
T [min]