Error Indication On The Analogue Output (Namur); Automatic Sensor Recovery (Arc) - E+E Elektronik EE1900 User Manual

Humidity module for climate chambers
Hide thumbs Also See for EE1900:
Table of Contents

Advertisement

4.3

Error Indication on the Analogue Output (NAMUR)

The EE1900 features an error indication on the analogue output according to NAMUR
recommendations. A device failure is indicated by 11 V in case of voltage output and by 21 mA
in case of current output.
Important:
Upon delivery the NAMUR error indication is disabled. It can be enabled with the free EE-PCS Product
Configuration Software.
5

Automatic Sensor ReCovery (ARC)

Certain chemical pollution may lead to drift of the sensing element. Using the sensor recovery function,
the contamination is removed by a controlled heating of the sensor. After the heating process, the sensor
quickly returns to normal measuring conditions (refer to Fig. 5).
Start:
The ARC can be started either manually by pressing the "Sensor Recovery" button (No. 6 - Fig. 3) for
min. 3 seconds or automatically by applying a voltage of 5 - 30 V DC to the potential free input (No. 3
"Heat IN" - Fig. 3) for at least 3 seconds.
Abortion:
To cancel the ARC process, the same procedure is valid as to starting it: pressing the "Sensor
Recovery" button (No. 6 - Fig. 3) or a voltage of 5 - 30 V DC at the "Heat IN" interface for min. 3
seconds.
Duration:
A sensor recovery cycle takes around 23 minutes, of which 20 minutes heat-up time, followed by 3
minutes cooling time. The cooling time is necessary also in case the ARC process is cancelled.
in
out
Galvanically separated input/output
Fig. 4
ARC trigger and status connection
Important:
ARC can only be started if RH < 85%.
User Manual EE1900 Humidity Module for Climate Chambers
+
Power supply
5 - 30 V DC
GND
GND
ARCStatus
Pulldown: ≤100 k
GND
+
Power supply
5 - 30 V DC
GND
Current consumption in ARC mode
Supply
15 V DC
24 V DC
V
35 V DC
17 V AC
24 V AC
29 V AC
Consumption
~120 mA (max.)
~80 mA (max.)
~55 mA (max.)
~210 mA
(max.)
rms
~160 mA
(max.)
rms
~140 mA
(max.)
rms