Page 2
List of Revisions: Rev. No. Rev. Date List of Effective Pages: Page 1 of 18 August 2012 Page 7 of 18 August 2012 Page 13 of 18 August 2012 Page 2 of 18 August 2012 Page 8 of 18 August 2012 Page 14 of 18 August 2012 Page 3 of 18...
The output provides a voltage signal proportional to the acceleration. It is powered by a constant current line drive supply on the same wire as the output signal. Accelerometer monoaxial 4511-W-006 Fig. 1 Dimensions : All dimensions in millimeters Fig.
Page 5
The accelerometer is intended to measure the vibration of the rotor for HUMS purposes on three axes: X, Y and Z. It consists of a triaxial piezoelectric element and an electronic charge converter enclosed in a hermetic housing. The output provides a voltage signal proportional to the acceleration for each axis.
Technical Data 1.2.1 Specification – Type 4511-W-006 – Monoaxial General Characteristics Accelerometer Type IEPE – monoaxial Voltage Sensitivity (at 159.2 Hz) 10 mV/g Sensitivity Tolerance ±5% on the whole frequency range Measuring Range ± 500 g pk Output Full Scale Voltage ±5 V...
Page 7
Insulation Resistance >100 MΩ Dielectric Strength 500 Vdc Signal Ground Isolated from housing Connection Through a bifilar cable (double-shielded recommended) Single-shielded possible if connected to structure in both ends Environmental Characteristics Temperature Normal Operational Range – 51°C to +150°C Temperature Survival Range –...
1.2.2 Specification – Type 4524-W-001 - Triaxial General Characteristics Accelerometer Type IEPE – triaxial Voltage Sensitivity (at 159.2 Hz) 100 mV/g Sensitivity Tolerance ±5% on the whole frequency range Measuring Range ±50 g pk Output Full Scale Voltage ±5 V Amplitude Response ±5% from 2 Hz to 2 kHz Phase Response...
Page 9
Dielectric Strength 500 Vdc Signal Ground Isolated from housing Connection Through a quadrifilar cable (Double-shielded recommended. Single-shielded possible if connected to structure in both ends.) Environmental Characteristics Temperature Normal Operational Range – 45°C to +125°C Max operating sinusoidal vibration 50 g pk Max.
Ambient temperature: 25°C ±5°C Procedure 2.3.1 Appearance Check labelling. Check overall dimensions. Check electrical connections. Note: For Type 4511-W-006, see for specification of overall dimensions • Fig. 2 For Type 4524-W-001, see for specification of overall dimensions • Fig. 4 2.3.2...
Page 11
Mounting the transducer on the calibrator: Start the calibrator by pressing its button. The level displayed / recorded by the DMAU must be: Type 4511-W-006 – : 10 mV (accelerometer sensitivity × 10 ms × gain of DMAU) ± 8% •...
Process: Disconnect the cable of the transducer. Dismount the transducer. Connect the transducer to the CCLD Power Supply cable: For Type 4511-W-006 – , use CCLD Power Supply WB-1372 and the cable supplied with it. Monoaxial For Type 4524-W.001 - , use CCLD Power Supply WB-1453 and the cable supplied with it.
Section 3 Disassembly The transducer must not be disassembled or attempted to repair. If there is a problem, contact your Brüel & Kjær representative who may replace the transducer. BN1100 Revision: 0 Initial Issue: 21/08/2012 ATA 31-10-04 The document reference is online, please check the correspondence between the online documentation and the printed version.
Section 4 Standard and Special Tools Standard Tools Standard tools are used for removal/installation and connecting/disconnecting. 14 of 18 The document reference is online, please check the correspondence between the online documentation and the printed version.
Special Tools 4.2.1 References of special tools: The following special tools are available: Name Part Number WB -1453 Electrical testing of Type 4524-W -001 CCLD Power Supply WB -1372 Electrical testing of Type 4511-W -006 –W L3429/AO0526/JJ81 (included with Connect W B-1453 to Type 4524-W -001 CCLD Power Supply) not shown Cable...
4.2.2 Control of special tools : -Manufacturer control: CCLD Power Supply WB-1372 and WB-1453 and Calibration Exciter Type 4294 must be sent to Brüel & Kjaer every two (2) years for certification of their specifications. -On site Control of the CCLD power Supply WB1372 and WB1453: Press the switch in battery position.
Section 5 Consumables 9 V batteries 6 LR61 for CCLD Power Supply WB-1372, WB-1453 and Calibration Exciter Type 4294. BN1100 Revision: 0 Initial Issue: 21/08/2012 ATA 31-10-04 The document reference is online, please check the correspondence between the online documentation and the printed version.
Section 6 Cleaning, Storage and Transportation Clean the accelerometer with a dry smooth cloth. Store the equipment in a polythene bag, then in a cardboard box. When the unit is packed as above, it must be stored in clean, dry, well-ventilated premises free from corrosive fumes. The temperature should be between –40 to +70°C and relative humidity between 20% and 85%.