Product specifications

Type:
Measuring range:
Resolution:
Laboratory Accuracy:
Display:
Acoustic:
Operating temperature:
Automatic Switch-off:
Weight:
Long term storage ranges
Storage/transport temperature:
Humidity:
Battery:
Pressure:
This infrared thermometer meets requirements established in ASTM Standard E 1965-98 (for the
thermometer system) . Full responsibility for the conformance of the product to the standard is
assumed by Kaz USA, Inc ., a Helen of Troy Company, Health & Home, 400 Donald Lynch Blvd ., Suite 300,
Marlborough, MA 01752
ASTM laboratory accuracy requirements for the thermometer only in the display range of 36 °C to 39°C
(96 .8 °F to 102 .2 °F) for infrared thermometers is ±0 .3 °C (±0 .5 °F), whereas for mercury-in-glass and
electronic thermometers, the requirement per ASTM Standards E 667-86 and E 1112-86 is ±0 .1 °C (±0 .2 °F) .
This device conforms to the following standards:
EN 60601-1:2014 Medical electrical equipment . General requirements for basic safety and
essential performance
ASTM E1965-98:2016 - Standard Specification for Infrared Thermometers for Intermittent Determination
of Patient Temperature
ISO 80601-2-56:2017 Particular requirements for basic safety and essential performance of clinical
thermometers for body temperature measurement
IEC 60601-1-2:2014: Medical electrical equipment – Part 1-2: General requirements for basic safety and
essential performance – Collateral Standard: Electromagnetic disturbances – Requirement and tests
NOTE: Do not use this device in the presence of electromagnetic or other interference outside the normal
range specified in IEC 60601-1-2:2007 .
EN 1041: 2008 Information supplied by the manufacturer of medical devices
IEC 60601-1-11: 2015 Medical electrical equipment – Part 1-11: General requirements for basic safety and
essential performance – Collateral standard: Requirements for medical electrical equipment and medical
electrical systems used in the home healthcare environment
No touch & touch forehead thermometer (BNT 400)
93 .2 °F – 109 .4 °F (34 °C – 43 °C)
0 .1 °F (0 .1 °C)
0 .4 °F for 95 °F – 107 .6 °F (0 .2 °C for the range 35 °C – 42 °C)
outside that range ±0 .5°F/0 .3°C
(Ambient Temperature: 59 °F to 104 °F (15 °C to 40 °C)
This thermometer displays a calculated oral equivalent estimate .
Liquid Crystal Display, 4 digits plus special icons
Audio:
Normal temp range = Green temp range: 1 long beep for 0 . 4 second duration
Fever = Red or Yellow temp range: 10 short beeps for 0 . 2 second duration
59 °F - 104 °F ( 15°C - 40°C )
Approx . 30 seconds after last measurement has been taken
3 .54 oz . (with batteries), 2 .91 oz . (w/o batteries)
-13 °F – 131 °F (-25 °C to 55 °C)
15–95% non-condensing
(2) AAA Batteries - at least 500 measurements
700-1060 hPA (0 .7-1 .06 atm)
13