Maintenance
SensoGate WA 133
Servicing the Drive Unit
The drive unit must be removed, for example:
• for general maintenance or inspection
• to clean the calibration chamber, e.g., after a sensor has broken
• to change the sensor / calibration-chamber gaskets
• in the event of a technical fault of the drive unit
Lubricants, O-Rings
For retractable fittings used in the chemical industry, the lubricant Syntheso Glep1 (silicone-free) is
applied. For retractable fittings used in the pharmaceutical / food industry (when FDA conformity
is required), the lubricant Beruglide L (silicone-free) is applied (registered according to NSF-H1).
On request, the lubricant Paraliq GTE 703 can be applied (good lubricating properties also at
increased temperatures and for a large number of stroke movements). This lubricant contains
silicone and is only used as special application on specific request.
Application
Lubricant
Materials of
elastomeric gaskets
FKM
FFKM
EPDM
FKM - FDA
FFKM - FDA
EPDM - FDA
+ means: suitable; – means: not suitable
Selected Cleaning Agents for Specific Applications
Application
Deposits and dirt
Limescale
Fat
Alcohol solubles
Proteins
Observe the chemical resistance of the retractable fitting!
1)
Pharma / Food
Beruglide L
(silicone-free)
FDA compliant
NSF-H1 registered
–
–
–
+
+
+
Cleaning agent
Water + brush
Dilute acids
Surfactant
Dilute alkali
Solvant
Pepsin/hydrochloric acid solution -
Paraliq GTE 703
(containing silicone)
FDA compliant
(USDA H1)
–
–
–
+
+
+
Specification
cold or hot
e.g., hydrochloric acid or sulfamic acid
-
e.g., sodium hydroxide
e.g., ethanol or isopropyl alcohol
37
Chemistry / Wastewater
Syntheso Glep 1
(silicone-free)
+
+
+
–
–
–
1)