Advertisement

General Purpose Aquabaths
057-533-00 • 7/21/05
Digital Water Baths:
18002, 18002-1CE
18007, 18007-1CE
18012, 18012-1CE
18022, 18022-1CE
18052, 18052-1CE
18102, 18102-1CE
18802, 18802-1CE
18902, 18902-1CE
Teflon® Coated:
18202, 18202-1CE
18212, 18212-1CE
Model No.
Basic Water Baths:
18000, 18000-1
18005, 18005-1
18010, 18010-1
18020, 18020-1
18050, 18050-1
18100, 18100-1
18800, 18800-1
18900, 18900-1

Advertisement

Table of Contents
loading

Summary of Contents for Barnstead International General Purpose Aquabath 18002

  • Page 1 General Purpose Aquabaths ™ Model No. Digital Water Baths: Basic Water Baths: 18002, 18002-1CE 18000, 18000-1 18007, 18007-1CE 18005, 18005-1 18012, 18012-1CE 18010, 18010-1 18022, 18022-1CE 18020, 18020-1 18052, 18052-1CE 18050, 18050-1 18102, 18102-1CE 18100, 18100-1 18802, 18802-1CE 18800, 18800-1 18902, 18902-1CE 18900, 18900-1 Teflon®...
  • Page 2: Table Of Contents

    Table of Contents Safety Information ................................3 Alert Signals ................................3 Description..................................4 Specifications ................................5 Digital Water Baths ..............................5 Electrical Requirements ..........................5 Temperature Range ............................5 Uniformity ..............................5 Control ................................5 Basic Water Baths ..............................5 Electrical Requirements ..........................5 Temperature Range ............................5 Uniformity ..............................5 Control ................................
  • Page 3: Safety Information

    Safety Information Read operation manual before using equipment to famil- Alert Signals iarize yourself the controls and accessories. When operating this equipment, always wear protective clothing, glasses and other accessories as specified by Warning the safety regulations of your organization. Warnings alert you to a possibility of personal injury.
  • Page 4: Description

    Description The Barnstead|Lab-Line General Purpose Aquabaths are designed for use in clinical research labs, pharmaceutical labs and a multitude of chemical, biological, research and industrial laboratory applications. They are ideal for sero- logical studies and numerous other applications requiring accurate temperature control. Special heaters positioned beneath the bath chamber transfer heat efficiently.
  • Page 5: Specifications

    Specifications Digital Water Baths Basic Water Baths Electrical Requirements Electrical Requirements 18002: 120V 50/60 Hz 500W 18000: 120V 50/60 Hz 500W 18002-1CE: 240V 50/60 Hz 500W 18000-1: 240V 50/60 Hz 500W 18007: 120V 50/60 Hz 1000W 18005: 120V 50/60 Hz 1000W 18007-1CE: 240V...
  • Page 6: Tank Capacity And Dimensions Digital Water Baths

    PECIFICATIONS Tank Capacity and Dimensions Digital Water Baths (applies to both 120V and 240V units) Model Liters Interior (cm) Exterior (cm) 18002 & 5-7/8" W x 11-5/8" D x 5" H 8-3/8" W x 14" D x 9" H 18002-1CE: (15 x 30 x 13) (21 x 36 x 23) 18007 &...
  • Page 7: Tank Capacity And Dimensions Basic Water Baths

    PECIFICATIONS Tank Capacity and Dimensions Basic Water Baths Model Liters Interior (cm) Exterior (cm) 18000 & 5-7/8” W x 11-5/8” D x 5” H 8-3/8” W x 14” D x 9” H 18000-1: (15 x 30 x 13) (21 x 36 x 23) 18005 &...
  • Page 8: Declaration Of Conformity

    PECIFICATIONS Declaration of Conformity (for 220-240 volt, CE models only) Barnstead International hereby declares under its sole responsibility that this product conforms with the tech- nical requirements of the following standards: EMC: EN 61000-3-2 Limits for harmonic current emissions EN 61000-3-3...
  • Page 9: Unpacking And Installation

    Unpacking and Installation Electrical Requirements 120 VAC models require a 120 VAC, 50/60 Hz power Note source. They are supplied with a 3-wire line cord. It Leave unit disconnected when not in should be plugged into an outlet designed for 3-prong use.
  • Page 10: Operation

    Operation Controls, Digital Water Baths Warning Do not use in the presence of flamma- ble or combustible materials or explo- sive gases. Fire or explosion may result causing death or severe injury. Temperature °C Power Warning Do not heat any substance above a temperature that will cause it to emit toxic fumes.
  • Page 11: Bath Temperature Is Maintained By Microprocessor Controller

    PERATION Bath Temperature is Maintained by Microprocessor Controller The control panel consists of a microprocessor temperature controller, LED temperature readout, power switch and one pilot light that indicate heater is operational. Setting the Temperature, Digital Water Baths • Turn the power switch ON. The display will flash the current revision number for about 5 seconds, then the temperature will be displayed.
  • Page 12: Controls-Basic Water Baths

    PERATION Controls-Basic Water Baths Control High Limit Control Panel Status Lamp (Amber) Control Thermostat Hi-Limit Status Lamp (Red) Water Bath Temperature is Regulated by Two Thermostats The control panel consists of two thermostats (control and high limit) and two pilot lights that indicate which thermo- stat is controlling temperature.
  • Page 13: Setting The Temperature, Basic Water Baths

    PERATION Setting the Temperature, Basic Water Baths • To set the bath temperature: • Turn the power switch ON. • Rotate the high limit knob clockwise to its maxi- mum setting. • Rotate the control thermostat to an approximate desired setting. Allow time for temperature of bath to stabilize.
  • Page 14: Maintenance

    Routine Tank Cleaning the distributor can provide, please call Clean the tank and components immersed in the bath the Barnstead International customer with a solution of mild soapy water. Use a sponge, cloth service department at 1-800-553- or synthetic sponge scrubber. Rinse well. Dry thoroughly.
  • Page 15: Some Stainless Guidelines To Consider

    AINTENANCE Some Stainless Guidelines to Consider Distilled water is recommended. Please note, if this water is very pure it may be corrosive to stainless. When filling a bath or incubator, ALWAYS ADD 2 to 40 PPM (20 TO 40 MG/LITER) DISODIUM PHOSPHATE OR SODIUM BICARBONATE, ADJUSTING DOSAGE TO PROVIDE A pH VALUE OF 7 TO 9.
  • Page 16: Cleansing Agents

    AINTENANCE BE ADVISED: NEVER USE THE FOLLOWING ON STAINLESS STEEL: Caution Aqua regia Extreme care must be taken when Ferric chloride handling these materials. Always Iodine work in an area with adequate ventila- Sodium acid tion. Use the precautions as outlined Sodium azide in the material safety data sheet (MSDS) and the manufacturer's...
  • Page 17: Materials Effective In Disinfecting

    • Alcohol BE ADVISED: this information is intended as guidelines only and Barnstead International makes no claim as to the suitability to any particular situation. Consult your staff chemist to determine what would be best for your stainless steel product and laboratory.
  • Page 18: Troubleshooting

    5. Once the thermostat is reset, reposition insulation, reinstall the bottom panel, plug the power cord back into the outlet and verify the unit is heating again. 6. If the water bath still fails to heat, contact Barnstead International customer service for further assistance.
  • Page 19: Replacement Parts

    Replacement Parts Description Part Number Cordset: 470-105-00 Grommet: 790-247-00 Diffuser Trays: 18022: 593-115-00 18052: 593-114-00 18802 (small), 18002: 593-110-00 18802 (large), 18007: 593-111-00 18102: 593-118-00 18012: 018-824-00 18900, 18902: 584-112-00 18202: 593-111-01 18212: 593-118-01 18020: 592-842-00 18050: 592-843-00 18800, 18000: 592-844-00 18800, 18005: 592-845-00...
  • Page 20 EPLACEMENT ARTS Description Part Number Fuse Holder: 18012, 18012-1, 18012-1CE, 18802, 18802-1, 18802-1CE, 18010, 18010-1, 18010-1CE, 18800, 18800-1, 18800-1CE: 330-297-00 18002: 330-237-00 18000, 18005, 18005-1, 18020, 18050, 18100-1, 18900: 330-273-01 Thermometer Holders: 18010: 592-170-00 18000, 18005, 18020, 18050, 18100,18800, 18900: 592-106-00 Heaters: 120V Units:...
  • Page 21: Ordering Procedures

    All parts listed herein may be ordered from the Barnstead International dealer from whom you pur- chased this unit or can be obtained promptly from the factory. When service or replacement parts are needed we ask that you check first with your dealer.
  • Page 24: Two Year Limited Warranty

    SALE, USE OR PERFORMANCE OF ANY PRODUCTS, WHETHER SUCH CLAIM IS BASED ON CON- TRACT, TORT (INCLUDING NEGLIGENCE), ANY THEORY OF STRICT LIABILITY OR REGULATORY ACTION. The name of the authorized Barnstead International dealer nearest you may be obtained by calling 1-800- 446-6060 (563-556-2241) or writing to: 2555 Kerper Boulevard...