Radiation Output Rate; Test Objective; Frequency; Required Equipment - Konica Minolta CS-3 Quality Control Manual

Xpress digital mammography
Hide thumbs Also See for CS-3:
Table of Contents

Advertisement

3 .5

Radiation Output Rate

3 .5 .1

Test Objective

To measure the radiation output rate of the equipment .
3 .5 .2

Frequency

• Initially at installation of the mammography X-ray unit
• ANNUALLY
• As part of an MEE
3 .5 .3

Required Equipment

• Calibrated dose meter
• 4cm of PMMA (acrylic) phantom
3 .5 .4

Procedure

1
Set at 28kVp in the standard mammography mode (Mo/Mo) and expose long enough to
achieve 3 seconds .
2
Position the detector of the dose meter 4 .5cm above the breast support table and centered
4cm in from the chest wall .
3
Position the paddle to come in contact with the detector . (On some units, a cassette must be
placed in the cassette holder to enable exposure .)
HINT
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
• Do not change the position of the detector during the measurements .
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
77