11
Exit the Analysis Tool .
• Select [File (F)], and then select the [Exit(X)] from the sub menu . The Analysis Tool screen is closed .
IMPORTANT
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
• Maintain a consistent exposure technique at all times . Use an AEC mode that varies only mAs, and does not allow variations
in kVp, target and filter material.
• When performing this test, always wait the same length of time between exposure and processing of the plate . Always erase
the REGIUS plates before running this test . Each mammography X-ray unit will produce a unique CNR baseline value .
Compare weekly test results only against the baseline values of the same unit .
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
HINT
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
• When establishing the baseline CNR value, repeat the process, from Step 5 through Step 11, four more times . Calculate the
average of the five CNR values and +/- 20% values, and record them in the Table RT-1 (Baseline CNR Value).
• The CNR value can be calculated by using a measurement tool of a diagnostic review workstation, instead of using the User
tool of REGIUS Console CS-3/CS-7 . In that case, process the cassettes under the CNR-WS exam tag and send the image
to the diagnostic review workstation .
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
2 .6 .5

Procedure (for CS-7)

1
Place PMMA (acrylic) phantom of thickness 40 mm on the breast support table .
2
Position the compression paddle as close as possible to the phantom surface .
3
Move the AEC sensor to its most forward position (closest to the chest side) . Make an expo-
sure with 28kVp in AEC mode using Mo target .
2 .6 Contrast to Noise Ratio (CNR) Check
33