IEM eConnect Instruction To Use page 23

Table of Contents

Advertisement

Electromagnetic immunity
The eConnect is designed for operation in an electromagnetic environment as specified below. Only use the
eConnect in such an environment.
Measurement of
interference emissions
Electrostatic discharge
(ESD) in accordance with
IEC 61000-4-2
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-11
Magnetic field at supply
frequency (50/60 Hz) in
accordance with IEC 61000-
4-8
Radiated RF interference
in accordance with IEC
61000-4-3
IEC 60601-1 test level
±/- 8kV contact discharge
±/- 15kV air discharge
30 A/m
380 - 390 MHz
27 V/m; PM 50%; 18 Hz
430 - 470 MHz
28 V/m; (FM ±5 kHz, 1
kHz Sinus)
PM; 18 Hz
Manufacturer's EMC guidelines
Compliance level
±/- 8kV contact discharge
±/- 15kV air discharge
Not applicable
Not applicable
Not applicable
30 A/m
380 - 390 MHz
27 V/m; PM 50%; 18 Hz
430 - 470 MHz
28 V/m; (FM ±5 kHz, 1
kHz Sinus)
PM; 18 Hz
Electromagnetic
environment – guidelines
Floors should consist of
wood or cement or ceramic
tiles. If the floor consists of
synthetic materials, relative
humidity must be at least
30%.
Magnetic fields at mains
frequency should match the
typical values found in
business and hospital
environments.
51
EN

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the eConnect and is the answer not in the manual?

Related Products for IEM eConnect

Table of Contents