6. HYGIENE TREATMENT

6.1 Cleaning

6.2 Disinfecting

74 •
WARNING!
Electric shock
The device is not de-energized when the on/off
key is pressed and the display goes out. Use of
fluids on the device may cause electric shock.
Ensure that the device is switched off before
performing any hygiene treatment.
Disconnect the power supply connector be-
fore performing any hygiene treatment.
Ensure that no fluids penetrate the device.
CAUTION!
Damage to device
Inappropriate detergents and disinfectants may
damage the sensitive surfaces of the device.
Use only disinfectants free of chlorine and al-
cohol which are explicitly suitable for acrylic
sheet and other sensitive surfaces (active in-
gredient: quaternary ammonium com-
pounds, for example).
Do not use caustic or abrasive detergents.
Do not use spirits or benzine.
Use a soft cloth dampened with mild soapsuds to
clean the surfaces of the device.
1. Check that your disinfectant is suitable for sensitive
surfaces and acrylic sheet.
2. Follow the instructions on the disinfectant.