Isolation Voltage
Environmental
Safety
Note
For UL and other safety certifications, refer to the product label or the
Product Certification
Electromagnetic Compatibility
-
1
The SCC
AI does not provide isolation between the two channels of the module. For more information, refer
to the
Connecting the Input Signals
© National Instruments Corporation
Channel-to-earth isolation
Continuous ...................................... 60 VDC,
Withstand ........................................ 2300 V
Channel-to-channel ................................ Nonisolated
Operating temperature............................ 0 to 50 °C
Storage temperature ............................... –20 to 70 °C
Humidity ................................................ 10 to 90% RH, noncondensing
Maximum altitude .................................. 2,000 m
Pollution Degree (indoor use only) ........ 2
This product meets the requirements of the following standards of safety
for electrical equipment for measurement, control, and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
section.
This product meets the requirements of the following EMC standards for
electrical equipment for measurement, control, and laboratory use:
•
EN 61326 (IEC 61326): Class A emissions; Basic immunity
•
EN 55011 (CISPR 11): Group 1, Class A emissions
•
AS/NZS CISPR 11: Group 1, Class A emissions
•
FCC 47 CFR Part 15B: Class A emissions
•
ICES-001: Class A emissions
section.
Measurement Category I
dielectric withstand type test
11
SCC-AI Series Isolated Analog Input Modules User Guide
verified by a 5 s
rms
1
Online