INTRODUCTION 1. Congratulations!! Congratulations..........3 Thank you for purchasing TPI products. The 665 is easy Product Description ..........3 to use and is built to last. It is backed by a 3 year limit- ed warranty. Please remember to complete and return SAFETY CONSIDERATIONS..........4...
B. SAFETY CONSIDERATIONS C. TECHNICAL DATA 1. Features and Benefits WARNING: Please follow manufacturers test procedures whenever possible. Do not attempt to Display 5 digit dual line with backlight measure unknown pressures. Damage to the instru- ment may result. Units of Seven units of measure (bar, kPa, GENERAL GUIDELINES Measure...
Page 4
B. SAFETY CONSIDERATIONS C. TECHNICAL DATA 1. Features and Benefits WARNING: Please follow manufacturers test procedures whenever possible. Do not attempt to Display 5 digit dual line with backlight measure unknown pressures. Damage to the instru- ment may result. Units of Seven units of measure (bar, kPa, GENERAL GUIDELINES Measure...
Press to turn the 665 on. 2. Zero the 665 by pressing 3. Connect the device under test to the “+” port of the 665. If a Press to activate the display backlight. Backlight differential measurement is being made, connect the other powers off after 15 seconds.
Page 8
Press to turn the 665 on. 2. Zero the 665 by pressing 3. Connect the device under test to the “+” port of the 665. If a Press to activate the display backlight. Backlight differential measurement is being made, connect the other powers off after 15 seconds.
Press to print readings to the optional A740 printer. and seconds. The 665 will display a 10 second countdown giving you time to 10. Exit “EDIT” mode by pressing align the back holes approximately 4” away from the printer win- dow.
Page 10
Press to print readings to the optional A740 printer. and seconds. The 665 will display a 10 second countdown giving you time to 10. Exit “EDIT” mode by pressing align the back holes approximately 4” away from the printer win- dow.
Page 11
The interval can be set from 1 second to 24 hours. can also be activated manually (See page 12). The 665 can log a total of 5000 data points. NOTE: The option- al A665 docking station and software is required to view logged 1.
The interval can be set from 1 second to 24 hours. can also be activated manually (See page 12). The 665 can log a total of 5000 data points. NOTE: The option- al A665 docking station and software is required to view logged 1.
Page 13
1. Activate “EDIT” mode by pressing 1. The logging interval and start time should already be set and the 665 should be displaying “L-St”. If not, follow steps 1 through 7 on page 12 and steps 1 through 6 on page 13 six times until “L-res”...
Page 14
1. Activate “EDIT” mode by pressing 1. The logging interval and start time should already be set and the 665 should be displaying “L-St”. If not, follow steps 1 through 7 on page 12 and steps 1 through 6 on page 13 six times until “L-res”...
4. Remove the batteries and observing polarity replace them with fresh batteries. 5. Re-install the battery holder and tighten the screws. To obtain service for your 665 please return it to: Cleaning TPI / Attn: Service 9615 SW Allen Blvd Suite 104...
Page 16
4. Remove the batteries and observing polarity replace them with fresh batteries. 5. Re-install the battery holder and tighten the screws. To obtain service for your 665 please return it to: Cleaning Use a mild detergent and slightly damp cloth to clean the surfaces of the 665.
Need help?
Do you have a question about the 665 and is the answer not in the manual?
Questions and answers