Programming menu

Below is a list summarising the device programming functions available. For more details, see from page
15.
To access programming press ENTER for several seconds, the green LED remains lit and the first LED on
the LED bar lights up (function F1).
To scroll through the menu of the programming functions, use the ENTER key.
To return to function F1 go to function F10 and press ENTER+SET.
To exit programming go to function F10 and hold down ENTER for several seconds. If the parameters were
saved correctly, the LED bar lights scroll, otherwise it begins flashing until you press SET.
Function
Bar
number
position
F1
1
F2
2
F3
3
F4
4
F5
5
F6
6
F7
7
F8
8
F9
9
F10
10
Rev.12/04/11
Description
Regulated output type selection
Minimum distance calibration
Middle diameter calibration
Maximum distance calibration
Setting 1
alarm threshold
st
Setting 2
alarm threshold
nd
Regulated output inversion
Led bar mode
Thermal probe exclusion
Exit menu saving the data
Keys
Value range
SET
0→1
SET
150mm→D.max -1
See detailed description of field calibration
(page 16)
SET
D.min.+1→1800mm
SET
150mm→1800mm
SET
150mm→1800mm
SET
0→1
SET
0→1
SET
0→1
ENTER
US3
Factory
setting
0
150mm
1800mm
1600mm
1700mm
0
0
1
-
-