The general safety requirements must be strictly adhered to when operating the device. The fact contained especially safety are observed. If you experience any further questions, please contact our technical support by phone or mail to tech@tempsens.com 1.2 Liability and warranty...
Before you Start 2.1 Safety Information Use the instrument only as specified in this manual. Otherwise, the protection provided by the instrument may be impaired. Refer to the safety information below and throughout the manual. The following definitions apply to the terms “Warning” and “Caution”. ...
Components and heater lifetime can be shortened by continuous high temperature operation. D o no t ch a n g e th e va l ue s o f th e c al i b ra ti o n c o n s t a n t s f r o m t h e f a c to r y s e t val u e s.
P: PHASE= RED WIRE N: NEUTRAL = BLACK WIRE E: EARTH =BLUE WIRE 4. Overview 4.1 Introduction: The ‘CALsys 1500BB’ provides a stable temperature source for laboratory calibration of radiation thermometers by comparison method up to a temperature range of 1500°C. The ‘CALsys 1500BB’...
Page 7
Silicon Carbide Cavity Open back panel sheet of CALsys 1500 and remove all Packing sheet from inside. Insert the 6 nos. silicon heater at back side of the inner chamber. As shown in below image. Insert the heaters in to the groves of SS chamber as per specified same sr.
Page 8
Fix the 6 heater into the chamber and Connect all the aluminum strip in series. Heating Element connection. connect the Phase and Netural wire as per given picture. Open the front plate of CALsys 1500BB and remove all packing sheets from inside. Open holding ceramic clamp as shown in figure.
Page 9
Insert cavity and ensure that the tail of cavity fix to the groove on the back panel. And make sure cavity properly insert at back panel hole as shown in image. After insert the cavity tight the holding cap of SS chamber by screw driver as per shown picture.
5. Operating Instruction (1) PID Controller (2) ON/OFF switch of controller (3) Power Cable (4) RS-232 (5) MCB (for Heater) Safety controller for over heating temperature. (7) 1."R" type thermocouple Duplex for controlling and over temperature protection. Inner chamber of calsys 1500BB 5.1 Power Plug the black body power cord into a mains outlet of the proper voltage, frequency and current capability.
Note: All other controller parameters are company set and locked. It is recommended not to change them. When the source is operated at any temperature above ambient, the front face and plate become hot. Down will switch the CALsys 1500BB ON or OFF to raise or lower the temperature. The displayed temperature will gradually change until it reaches the setpoint temperature.
6. Operation of Controller FRONT PANEL LAYOUT Operator Buttons 6.1 The Temperature Controller The controller has a dual display, the upper display indicates the measured temperature, and the lower display indicates the desired temperature or set point. 6.2 Altering the Set point To change the set point of the controller use the UP and DOWN keys to raise and lower the set point to the required value.
7.2 Trouble shooting 1. Unit fails to operate Check fuse if it is tripped switch is ON. If not power ON of calsys 1500 BB consult us. 2. Unit unstable Controller parameter has been interfered, consult us.
Communication can be done by clicking on connect and select correct COM port address (fig. 1) where furnace is connected. Also user has to select type of controller...
Page 15
1.2.2 Scale Trend: -in scale trend you have to change the Y-Axis Min value 0 and Y-AxisMax value 1500.MaximumTime Span, Minutes have to save data the data logging up to 120 minutes, than clickon start graph button. After complete the task clickon save to file button. Set Point (°C): - in which you can set temperature of furnace as your requirement.
Page 16
TEMPSENS reserves the right to make changes, without further notice, to any product herein to improve reliability, function, or design . TEMPSENS does not assume any liability arising out of the application or use of any product described herein, neither does it convey any license under its patent right nor the right of others.
Need help?
Do you have a question about the CALsys 1500 BB and is the answer not in the manual?
Questions and answers