Specifications; Troubleshooting; Warranty - Welch Allyn 48700 User Manual

Halogen exam light iii
Hide thumbs Also See for 48700:
Table of Contents

Advertisement

Available languages

Available languages

Mobile Stand
(Model # 48850)
Caster/Base Assembly
 
Instructions
:
Push casters into the holes of
the base until they "snap" into
Base
place. Alternate casters as
shown.
Base/Column Assembly
 
Instructions
:
Locking Caster
1.
Insert column into
center hole of base.
Locking
 
2.
Insert .25 inch (.64 cm)
Caster
lock washer, .75 inch
(1.9 cm) flat washer and
1 1/4" (3.2 cm) socket head bolt into center hole. Tighten socket head
bolt with hex socket wrench.
.75 inch (1.9 cm)
Flat Washer
.25 inch (.64 cm)
Lock Washer
Mounting Plate
Socket Head Bolt
Column
Hex Socket Wrench
 
(Allen Wrench)
Base
 
Exam Light/Mounting Plate Instructions
:
1.
Place the keyhole pins on the back of the Exam Light into the slots on
the top of the mounting plate and slide down.
2.
Line up the screw holes on the bottom of the Exam Light with the screw
holes on the mounting plate.
3.
Insert the .50 inch (1.27 cm) screws through the bottom of the mounting
plate and tighten with a phillips head screw driver (not included).
Exam Light
Keyhole Pin in slot.
.50 inch (1.27 cm)
mounting screw
 
SET-UP AND OPERATION
Set Up
1.
Secure the Light Box to one of the mounting options per the appropriate
instructions.
2.
Insert the Fiberoptic Light Pipe, Fiberoptic Light Cable, or Neonate Light
Pipe into the light pipe receptacle on the top of the Light Box.
3.
Plug the power cord into the base of the Light Box and the remaining
end into a grounded AC outlet.
4.
Toggle the light switch to the "ON" position to operate.
NOTE: Remember that the Fiberoptic Light Pipe, Fiberoptic Light Cable, and
Neonate Light Pipe contain several thousand individual optical fibers,
therefore appropriate handling is required to ensure long fiber life. Do not
bend the flexible portions of the light pipe at acute angles as this will break
individual fibers and reduce light transmission. For example, do not bend the
semi-flexible middle portion of the Fiberoptic Light Pipe more than 180
degrees and do not bend the proximal and distal sections more than 90
degrees. Be careful not to pull on the Fiberoptic Light Cable as this may cause
fibers to break.
Operation Of Light Pipe
The flexible portions of the Fiberoptic Light Pipe are designed to be
articulated, but not twisted. Twisting or applying torque to the flexible
sections will greatly decrease the life of the Light Pipe. This type of
movement will cause premature drooping of the Light Pipe (will not stay in
place once positioned) and eventually will result in breakage of the fiberoptic
bundle. For best results, always position the Light Pipe by moving it in the
same direction as gravity. For example, if the pipe needs to be raised, raise it
above the desired position and then lower it into place.
NOTE: Upon manufacture, the Light Pipe is tested to assure maintenance of
positioning and to verify the absence of noise upon articulation. Over a period
of time and use, it is normal for the Light Pipe positioning capability to reduce
and soften becoming less obedient. Abnormal softening or occurrence of
noise upon articulation should be reported to Welch Allyn Technical Service
(315-685-4560) immediately.
Operation of Focusing Sleeve
The focusing sleeve at the end of the Fiberoptic Light Pipe and Fiberoptic
Light Cable provides adjustment of the illumination spot size. Twisting the
knurled portion of the focusing sleeve in a clockwise direction will make the
spot smaller. Twisting the knurled portion of the focusing sleeve in a
counterclockwise direction will increase the spot size.
LAMP REPLACEMENT
CAUTION: Lamp may be hot. Allow the lamp to cool before replacing.
1.
Turn the unit off and unplug the power cord from both the wall outlet
and the Light Box.
2.
Remove the unit from the patient area.
3.
Using a screwdriver, loosen the single panel fastener that protrudes
from the bottom of the unit. As the screw is loosened, there will be a
 
noticeable "drop" as the front cover is released.
4.
Slightly lift the bottom of the front cover toward you and then slide the
 
cover up to remove it.
5.
For initial lamp replacement only: Grasp wire clip at top of lamp. Rotate
 
clip toward the base of the unit until it pops out.
Discard clip. It was
for shipping purposes only.
6.
 
Grasp the lamp connector and pull the lamp straight out of its holder.
7.
Pull the lamp from its connector and replace it with Welch Allyn
 
Replacement Lamp #04200.
8.
Push the lamp back into the holder.
9.
Attach the front cover by sliding the tabs at the top of the cover into the
slots at the top of the back housing and then tightening the screw at
the bottom of the unit.
10.
Return the unit to the patient area and plug power cord back into
bottom of unit and wall outlet.
 
 
 
 
 
FUSE REPLACEMENT
1.
Turn the unit off and remove the power cord from both the wall outlet
and the light box.
2.
Press the tabs on both ends of the fuse drawer in and pull the fuse
 
drawer out.
3.
Remove and replace new fuses of the same type. There is no required
 
orientation of the fuses. Be sure to replace both fuses.
4.
Reinsert the fuse holder by pressing it into the power connector until it
snaps into place.
5.
If the new fuses blow, immediately contact Welch Allyn Technical
Service (315-685-4560).
 
 
Fuse Tabs
CLEANING
Focusing Sleeve
For best results, use the disposable sheath model #52640.
1.
Remove the focusing sleeve from the Fiberoptic Light Pipe or Fiberoptic
Light Cable by rotating it counterclockwise to unthread.
2.
Use a cotton swab soaked in warm water and mild detergent to clean
the distal end of the Fiberoptic Light Pipe or Light Cable, the proximal
side of the Focusing Sleeve (small end), and distal window on the large
end on the Focusing Sleeve.
3.
Dry using a new cotton swab.
4.
Reassemble the Focusing Sleeve to the Light Pipe or Light Cable by
rotating clockwise.
WARNING: Do not clean any part of the focusing sleeve with alcohol. Clean
only with a mild detergent and water.
NOTE: The focusing sleeve should not be sterilized or immersed.
Fiberoptic Light Pipe, Light Cable, Neonate Light Pipe
For best results, use the disposable sheath model #52640.
External surfaces may be cleaned with a solution of warm water and mild
detergent.
NOTE: The Fiberoptic Light Pipe, Light Cable, and Neonate Light Pipe should
not be sterilized or immersed.
Light Box
Note: The Light Box should not be sterilized or immersed.
1.
Prior to cleaning, turn power off and disconnect the power cord from
both the wall outlet and the Light Box.
2.
The external surfaces may be cleaned by wiping with a cloth dampened
with mild soap and water or by using the following cleaning/
disinfecting solutions. DO NOT IMMERSE.
70% Isopropyl Alcohol
10% mild bleach solution
10% Iodine Based Solution
The power cord may be wiped clean using the same cleaning/disinfecting
solutions listed above. Wipe dry with a clean cloth. Care must be taken not to
get the prongs wet. DO NOT IMMERSE.

TROUBLESHOOTING

If lamp is not
illuminating
Possible Cause
Solution
and...
fan is not oper-
Light box is not
Plug the power cord into
ating.
plugged into either
the wall outlet and power
the wall outlet or the
cord socket at bottom of
power cord socket.
unit.
Power switch is not
Depress the green side of
turned on.
the power switch.
One or both fuses
Replace both fuses. Follow
have blown.
the instructions in the fuse
replacement section. If
fuses continue to blow,
return Light Box to an
authorized repair center.
Power cord is dam-
Replace the power cord
aged. This may cause
with a new Welch Allyn
intermittent opera-
power cord.
tion.
No power is sup-
Verify that the wall outlet
plied to the wall out-
has power by checking the
let.
circuit breaker status and
by operating other equip-
ment at that wall outlet.
fan is operating.
The lamp has burned
Replace with a new Welch
out.
Allyn lamp, model # 04200.
Refer to the Lamp Replace-
ment Section.
If light output
Lamp is not in opti-
Rotate lamp 90° in the
seems low.
mal orientation.
lamp holder.
MAINTENANCE AND REPAIR
For minor trouble, refer to the troubleshooting section in this manual for
possible causes and corrective action. Only qualified personnel should make
electrical inspections of the Welch Allyn Light Box. To locate qualified
personnel, contact your local authorized Welch Allyn distributor, or call Welch
Allyn directly at 315-685-4560. Tampering with the unit will automatically
void all warranties expressed or implied.

WARRANTY

Light Box - Welch Allyn guarantees that the Light Box is free of any
manufacturing defects or Welch Allyn will repair or replace, free of charge,
any parts proven to be defective through causes other than misuse, neglect,
damage in shipment and normal wear and tear. The warranty is for a one year
period from the original date of purchase.
Fiberoptic Light Pipe - The fiberoptic light pipe has been designed to not drift
to a new position once positioned for an exam if the Light Box is mounted
vertically according to the instructions. The warranty is for a one year period
from the original date of purchase provided that the light pipe has not been
tampered with or abused. Any Light Pipe that does not meet these
performance requirements will be repaired or replaced at no charge to the
customer.

SPECIFICATIONS

Electrical
Ref
Input
Output
Fuse
Model
48740
120VAC∼60Hz,
12VAC,
T500mAL,
400mA
3.0A
250V
48742
230VAC∼50Hz,
12VAC,
T315mAL,
250mA
3.0A
250V
48744
230/
12VAC,
T315mAL,
240VAC∼50Hz,
3.0A
250V
250mA
48748
100VAC∼50/
12VAC,
T630mAL,
60Hz, 500mA
3.0A
250V
Power Cord
#18AWG, 3 wire grounded cord set, 8 feet (2.5 meters) long.
Physical Characteristics
Weight:
4.75 lb (2.16 Kg), light box only
Light Box
H: 8.75 inches (22.23 cm)
Dimensions:
W: 5.25 inches (13.34 cm)
D: 3.50 inches (8.89 cm)
Environment
Operating:
60°F (15°C) to 104°F (40°C), 75%
R.H. Max, 500 hPa - 1060 hPa Altitude
Transport/Storage:
-4°F (-20°C) to 120°F (49°C), 95%
R.H. Max, 500 hPa - 1060 hPa Altitude
Equipment Classification
Class I, Continuous Operation
IPXØ: Not protected against the ingress of water.
Approvals
IEC 60601-1, UL 2601-1, CAN/CSA C22.2 No
601.1, AS3200.1, IEC 60601-1-2
US
WELCH ALLYN AUSTRALIA PTY LTD
EMC Framework of Australia
18-20 ORION ROAD
LAND COVE, NSW 2066
AUSTRALIA
The CE mark on this product indicates that it has been
tested to and conforms with the provisions noted within
the 93/42/EEC Medical Device Directive.
Authorized European Representative Address:
European Regulatory Manager
Welch Allyn Ltd.,
Kells Road, Navan,
County Meath, Republic of Ireland.
Tel.: 353-46-79060
Fax: 353-46-27128
NOTE: This product complies with current required standards for
electromagnetic interference and should not present problems to other
equipment or be affected by other devices. As a precaution, avoid using this
device in close proximity to other equipment.
NOTE: This product contains no hazardous materials. Its disposal will not
contaminate or harm the environment, or present any risk to individuals
disposing of the product. As a precaution, it is recommended you contact your
local disposal and/or recycling authority for information regarding the
disposal of the equipment.