The Hybrid-Power The Hybrid-Power system combines the pressure redistribution properties using high specification foams with advanced alternating cell technology. With these 2 key features, we can offer the patient a true ‘step up, step down’ alternating therapy to repair and maintain healthy skin cells as well as providing a comfortable and stable platform.
Fire Warning The Hybrid-Power system complies with all relevant UK fire safety standards. In order to reduce the risk of fire: DO NOT SMOKE - Smoking will contaminate the product and is NOT permitted around or on the mattress system. This is a common cause of fatal fires. This system uses room air for circulation through the mattress.
IMPORTANT SAFEGUARDS READ ALL INSTRUCTIONS BEFORE USE DANGER - To reduce risk of electrocution: 1. Always unplug this product immediately after using. 2. Do not use while bathing. 3. Do not place or store product where it can fall or be pulled into a tub or sink. 4.
Page 6
Disposal of Electrical & Electronic Equipment (WEEE) Product conforms to all relevant standards (See Declaration of Conformity) 5 – Hybrid-Power...
Introduction Product description The Hybrid-Power mattress is constructed from a number of carefully selected and purpose built components using clinical testing to optimise patients comfort and support. The mattress will support a patient weight up to 254kg. Any patient weight under 30kg requires the carer to conduct a risk assessment.
Page 8
4. - High pressure cell: These cells have been specially designed and profiled to allow an increased air flow to lower the overall peak pressures around the heels. Cover A waterproof vapour permeable cover with high frequency welds which can be cleaned with higher concentrations of sodium hypochlorite without compromising the covers protective properties.
Installation Unpackaging When unpacking the Hybrid-Power System, check for any damage which may have occurred during transit. If found to be damaged remove the system from use and contact Drive DeVilbiss Healthcare Ltd. Setting up 1. Remove all existing mattress surfaces from the bed frame 2.
Pump Functions Keys Function and Description Patient Weight Settings 3 comfort levels are available for a variety of patient weights. Use the- / + to adjust the weight settings accordingly On/Off Standby Main power rocker switch can be located at the side of the pump. The standby button is located on the front panel Static All cells in the mattress will be inflated by pressing this button.
Storage For the ease of storage, mattresses can be stacked on top of each other. To prevent mattresses becoming contaminated and/or damaged during storage: Mattresses should be stored flat Mattresses should be stored within a polythene cover or bag Do not store other objects such as side rails or bed ends on top of mattresses Allow adequate space for the stored mattresses to be manoeuvred by two people Damage Prevention Take extra care not to puncture the cover as this will allow fluid ingress and ultimately contaminate...
Cleaning The following guidelines are suggested by Drive DeVilbiss Healthcare as being suitable infection control procedures. Further information is available upon request. Pump Unit General cleaning may be affected by using a cloth dampened with a mild detergent and water solution.
Page 13
Laundering Before laundering the mattress, covers should be completely removed. Where required, mattress covers can be laundered as follows: o Pre wash: 95°C + 15 minutes o Main wash: 95°C + 15 minutes This should be followed by a cold rinse and extraction. Drying Mattress inners and tube sets should be wiped with a clean and dry cloth before reinserting.
Mattresses should be checked regularly to ensure they remain ‘fit for purpose’, clinically effective and pose no risk of infection to either the patient or the carer. Drive DeVilbiss Healthcare Ltd. recommend that a thorough inspection of both the interior (foam core and cells) and exterior (cover) of the mattress is carried out on a weekly basis or, each time a new patient is placed on the mattress.
This is normal in alternating mode. Air outlets take turns to from some of the air produce air during their cycle time outlets of the tube If fault still continues, a Drive DeVilbiss Healthcare approved connectors engineer will be required to repair the fault Pump is noisy...
The addition of components or accessories may affect the EMI susceptibility of the device. Do not fit accessories other than Drive DeVilbiss Healthcare authorised accessories. 15 – Hybrid-Power...
Warranty Your Drive DeVilbiss Healthcare product is warranted to be free of defects in materials and workmanship for 5 years on mattresses and 2 year on pumps from date of purchase. This device was built to exacting standards and carefully inspected prior to shipment. This warranty is an expression of our confidence in the materials and workmanship of our products and our assurance to the customer.
Page 18
The Team at Drive DeVilbiss develops its products to give our customers the freedom to live independently. This encompasses their daily home life and provides them with the opportunity to enjoy an outing with family and friends. Our goal is to develop a range that will provide individuals with a chance to enjoy everyday life.