Error messages

LED display
»Start« LED fleshes orange,
LED for W&H Service Oil F1, MD-200
lit up orange
»Start« LED fleshes orange,
LED for W&H Activefluid, MC-1100
lit up orange
»Start« LED fleshes orange,
LED for W&H Activefluid, MC-1100
lit up orange
Description
W&H Service Oil F1, MD-200 cartridge
empty or not in the unit
Electronics defective
W&H Activefluid, MC-1100 cartridge empty
or not in the unit
Electronics defective
Air supply fails during the preparation
process
Remedy
> Replace or insert the cartridge.
> Press the »Start« button once.
> Carry out the initial filling (see page 27):
Press the »Start« button for approx. 5 seconds.
> Send the entire unit to an authorized W&H service
partner.
> Replace or insert the cartridge.
> Press the »Start« button once.
> Carry out the initial filling (see page 27):
Press the »Start« button for approx. 5 seconds.
> Send the entire unit to an authorized W&H service
partner.
> Re-establish the main air supply.
> Press the »Start« button once.
> Carry out the initial filling (see page 27):
Press the »Start« button for approx. 5 seconds.
45