4 Installation
4.4 Assemble the product
Install the Electrical Power
Supply kit
Before connecting Peristaltic Pump P-1 to the electrical power supply, read the following
instructions carefully:
Step
1
2
3
40
Action
Two Electrical Power Supply Kits are supplied with Peristaltic Pump P-1, one
for 100–120 V and one for 220–240 V. Choose the kit appropriate to your
mains supply voltage, and discard the other kit.
Remove the yellow warning label covering the fuse/voltage selector on the
rear panel, see
Electrical connections on the side panel, on page
Open the fuse/voltage selector with the key provided or with a thin screw-
driver (step I in the illustration).
I
II
III
Peristaltic Pump P-1 Operating Instructions 28962201 AF