Electromagnetic Environment-Guidance - CAIRE Spirit 300 User Manual

Hide thumbs Also See for Spirit 300:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Safety
WARNING: PORTABLE RF COMMUNICATIONS EQUIPMENT (INCLUDING PERIPHERALS SUCH
AS ANTENNA CABLES AND EXTERNAL ANTENNAS) SHOULD BE USED NO CLOSER THAN 30 CM
(12 INCHES) TO ANY PART OF THE SPIRIT, INCLUDING CABLES SPECIFIED BY THE MANUFACTURER.
OTHERWISE, DEGRADATION OF THE PERFORMANCE OF THIS EQUIPMENT COULD RESULT.
WARNING: USE OF ACCESSORIES, TRANSDUCERS AND CABLES OTHER THAN THOSE SPECIFIED OR PROVIDED
BY THE MANUFACTURER OF THIS EQUIPMENT COULD RESULT IN INCREASED ELECTROMAGNETIC EMISSIONS
OR DECREASED ELECTROMAGNETIC IMMUNITY OF THIS EQUIPMENT AND RESULT IN IMPROPER OPERATION.
WARNING: USE OF THIS EQUIPMENT ADJACENT TO OR STACKED WITH OTHER EQUIPMENT SHOULD BE
AVOIDED BECAUSE IT COULD RESULT IN IMPROPER OPERATION. IF SUCH USE IS NECESSARY, THIS EQUIPMENT
AND THE OTHER EQUIPMENT SHOULD BE OBSERVED TO VERIFY THAT THEY ARE OPERATING NORMALLY.
Caution: Medical Electrical Equipment needs special precautions regarding Electromagnetic
compatibility (EMC) and needs to be installed and put into service according to the EMC information
provided in this manual.
Caution: Portable and mobile radio frequency (RF) communications equipment can affect Medical
Electrical Equipment.
Caution: The Spirit should not be used adjacent to or stacked with other equipment and that if
adjacent or stacked use is necessary, the Spirit should be observed to verify normal operation in the
configuration in which it will be used.
Table 1
Guidance and Manufacturer's declaration—electromagnetic emissions
The
is intended for use in the electromagnetic environment specified below. The customer or the
Spirit
user of the
should assure that it is used in such an environment.
Spirit
Emissions test
RF emissions
CISPR 11
RF emissions CISPR 11 Class B
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Compliance
Electromagnetic environment—guidance
Group 1
The
Spirit
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The
Spirit
Not applicable domestic establishments and those directly connected to the
public low-voltage power supply network that supplies
Not applicable buildings used for domestic purposes.
uses RF energy only for internal function.
is suitable for use in all establishments, including
PN MN235-C4 C | User Manual
Spirit
15 - ENG

Advertisement

Table of Contents
loading

This manual is also suitable for:

Spirit 600Spirit 1200Spirit series

Table of Contents