Exception Message; Product Performance - Harobot HRB-E0110Q Instruction Manual

Non-contact forehead
Table of Contents

Advertisement

Harobot

[Exception message]

Fault description
Screen display
"L0 "or"Hi"
no response
No display or abnormal display
Silent tone
Shut down immediately after
power on

[Product performance]

Product model
Anthropometric range
Measurement accuracy
Display resolution
Type of protection against
electric shock
Degree of protection against
electric shock
Protection degree of liquid
Operation mode
AP / APG device or not
Electromagnetic
compatibility-[GB4824-2013]
Permanent equipment or not
working condition
Resolvent
1. Check the measurement object. Measurement cannot be guaranteed
under the conditions of forehead blowing, dripping, sweating, applying
cosmetics, etc.
2. Check the temperature offset setting, factory set to 0.0.
3. Check the operating environment. Environmental change has great
influence on measurement.
4. If the ambient temperature changes too much, or the ultra-high
temperature object is directly transferred to the low temperature object,
the measurement difference will occur. It should be used in a relatively
stable environment after 10 minutes of storage to obtain a new test heat
balance.
5. Check the measurement distance (3-5cm).
Reload the battery; check if it operates under the setup menu
Reload battery
Check if the Beep setting is off
Check the battery capacity and reload the battery
Memory
inlet
battery
HRB-E0110Q
32.0℃ ~ 42.5℃
Within 33.0℃ ~ 37.0℃,Not more than ±0.2℃
0.1℃
32 group
Internal power equipment
Type B application part
IPX0
Continuous operation
not
Group I B class
not
Temperature 1-50 ℃ relative humidity ≤ 85%
DC 3V(2 section AAA battery)