Memory Function - Harobot HRB-E0110Q Instruction Manual

Non-contact forehead
Table of Contents

Advertisement

Harobot
Measurement mode setting:
In the power on state, press the "set" key briefly to switch the measurement mode to body
temperature and calibration.
Temperature unit setting - F1
Press and hold the "set" key for 2 seconds, F1 will be displayed on the screen, the icon of Celsius
temperature will flash, press the "-" key to switch the "Fahrenheit temperature" unit, press the
"+" key to switch the "Celsius temperature".
Temperature alarm point setting - F2
Press and hold the "set" key for 2 seconds, the screen will display F1, press and hold the "set" key
for 1 time to enter F2, then press and hold the + "key to increase the temperature by 0.1 ℃,
press and hold the" - "key to decrease the temperature by 0.1 ℃. Note: the alarm default value is
38 ℃ (100.4 ℉).
Temperature offset setting - F3
Press and hold the "set" key for 2 seconds, F1 will be displayed on the screen, press and hold the
"set" key for 2 times to enter F3, then press and hold the + "key to increase the offset value by
0.1 ℃, press and hold the" - "key to decrease the offset value by 0.1 ℃.
Tone switch - F4
Press and hold the "set" key for 2 seconds, F1 will be displayed on the screen, press and hold the
"set" key for 3 times to enter F4, then press and hold the "+" key to start the buzzer, and press
and hold the "-" key to close the buzzer.
Tips:
1. After setting the parameters, you need to save the settings. If you exit during shutdown, the new
settings will not be saved.
2. After entering the setting menu, pressing the measurement key is invalid and the thermometer
will not measure.
3. Body temperature mode is used to measure human body temperature, which is obtained by
dynamic compensation of ambient temperature and forehead surface temperature.
4. According to the influence of environmental temperature, measurement distance, specific
individual skin difference and other factors, the temperature deviation value can be corrected by
the difference between the target measurement value and the real value, and the correction
range is limited to - 3.0 ℃ ~ 3.0 ℃. See the setting operation for specific settings. The factory
default value is 0.0 ℃.

[Memory function]

This product can automatically store 32 sets of latest measured values:
In the power on / off state, press "+" or "-" to enter the memory query.
Short press "+" or "-" to view the 1st and 2nd in turn Group 32 memory.