Technical Manual electronIc crane Scale : calIBratIon ModelS: ocS-M & ocS-l taBle of contentS dISPlaY IllUStractIon KeY fUnctIon In ParaMeter SettInG/calIBratIon ParaMeter SettInG calIBratIon troUBleSHootInG GUIde Hantverksvägen Tel. +46 (0)176-208920 Technical_Manual_OCS-M_V1 Scandinavia’s largest webshop for scales 76493 Vaddo, Sweden www.vetek.com...
Page 2
+ net weight exceed full capacity + 9e [ -lb- ] low battery warning turn off automatically one minute later [ U*.** ] the voltage of current battery [ end ] when parameter setting or calibration ready [ off ] turn off www.vetek.com info@vetek.com...
Page 3
[ - ], 3= stable, after 15 minutes, scale display [ - ],after 60 minutes, turn off automatically. [ →0← ] [ off=1 ] Parameter set in factory [ Mr ] [ end ] Parameter setting finished www.vetek.com info@vetek.com...
Page 4
[ end ] Calibration finished, return ATTENTION: Scale tare weight can’t exceed 50%f.S., the range between load1, load2, load3 can’t less than 20%F.S., if the range less than 20%F.S., calibration finished automatically and return. load1 < load2 < load3 www.vetek.com info@vetek.com...
Page 5
Battery can’t recharge Battery is damaged replace battery Battery can’t recharge outlet is damaged replace the outlet Remote distance shortened receiver window is too dirty check and clean Remote distance shortened controller has low battery replace controller battery www.vetek.com info@vetek.com...