Keysight Y1225A Installation Manual
Keysight Y1225A Installation Manual

Keysight Y1225A Installation Manual

Rack mount kit

Advertisement

Keysight Y1225A Rack Mount Kit
Installation
This installation note contains procedures for installing the Y1225A Rack Mount
Kit for the Keysight M9502A 2-slot AXIe Chassis in a standard EIA rack.

Parts List

The parts included with each rack mount adapter option and the support rail are
shown in the following table:

General Guidelines

In rack mounting the chassis, follow these guidelines:
1 Always begin installing the chassis and heavy instruments at the bottom of
2 Anti-tipping feet, if available with the rack, should always be extended.
3 The heaviest instrument of chassis should always be mounted at the bottom
4 For maximum cooling and optimum rack thermal efficiency, place the
5 If instruments or modules in the chassis are consuming maximum power, 1U
Quantity
Description
1
Rack Flange Kit
4
Rack Flange Screws M4x6 Flat Head
2
Support Rails
8
Channel Nuts (for support rail-to-rack)
4
Support Rail-to-Rack Screws - 10-32 x 0.5
4
Dress Adapter Screws for rack flanges
the rack and work up. This maintains a lower center of gravity and reduces the
possibility of the rack tipping.
of the rack.
instrument or chassis with the greatest power consumption towards the top of
the rack. This promotes efficient cooling since heat rises when placed nearer
to the top of the rack, high power instruments will not unnecessarily heat
other instruments. However, in doing this, do not violate the guideline that
the heaviest instruments should be placed at the bottom of the rack.
of rack space is required for ventilation below the chassis.
Part Number
N5650-00020
N/A
M9018-00001
0590-0804
2680-0278
E9900-06001

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Y1225A and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Keysight Y1225A

  • Page 1 Installation This installation note contains procedures for installing the Y1225A Rack Mount Kit for the Keysight M9502A 2-slot AXIe Chassis in a standard EIA rack. Parts List The parts included with each rack mount adapter option and the support rail are...
  • Page 2 Keysight Y1225A Rack Mount Kit Installation Procedure 1 The M9502A AXIe chassis has a single carry handle which must be removed before removing the bumpers. Remove the two #2 Phillips screws securing the handle.See Figure 1. Figure 1 . Removing the Carry Handle 2 Remove the four #2 Phillips screws from each side of both bumpers.
  • Page 3 Keysight Y1225A Rack Mount Kit Installation Rack Unit Center Hole Channel Rack Nuts Unit Figure 3 . Center Holes of EIA Rack, Attaching Channel Nuts 4 Attach the rack mount front handle flanges to the chassis. Refer to Figure 4.
  • Page 4 Keysight Y1225A Rack Mount Kit Installation Two Flathead Screws Figure 4 . Attaching Rack Mount Flanges to M9502A Chassis 5 Position the rails on the rack’s vertical columns where the chassis is to be mounted. Align the adapter holes over the “center” holes of the EIA rack units.
  • Page 5 Keysight Y1225A Rack Mount Kit Installation If the rack will be moved excessively or over long distances, you should use a strap to secure the chassis to the rack slides. 10-32 x 0.5 screws Figure 5 . Attaching the Support Rails to the Rack.
  • Page 6 This information is subject to change without notice. © Keysight Technologies 2014 Edition 2 September 2014 *Y1225-90001* Y1225-90001...

Table of Contents