5.6 Alarms
Alarm System On or Off
1. Turn on alarm system by pressing alarm button for about 5 seconds. When switching alarm on or
off, a short audible alarm will come out for notice and the pilot lamp of Alarm will either flash or not.
When alarm system is armed, alarm pilot lamp is no lighting. When Alarm system is disarmed, LED
LAMP of alarm is lighting to indicate to user to be noted.
2. When the alarm system is switched on, or after any values have been reprogrammed, the Alarm
system is inactive until the set point values (±1) are achieved and maintained for more than 3 minutes,
after which the Alarm System is armed. This is to activate alarm after stabilization of temp and CO₂
percentage.
Alarm
If an alarm occurs, a beep will be heard and a fast flashing will occur on the RUN lamp indicator.
- If temperature deviates more than ±1℃ from set point for more than 8 min. the alarm will occur.
-Pressing mute button once will temporarily deactivate (10 minute delay). If temperature is not
recovered in this delay time, alarm occurs again.
-Alarm will automatically stop once temperature is recovered into tolerance range (±1℃)
-If CO₂ deviate more than ± 1% from set point for 8 min., the alarm will occur.
-Alarm will automatically stop once temperature is recovered into tolerance range.
-Pressing mute button once will temporarily deactivate (10 minute delay). If temperature is not
recovered in this delay time, alarm occurs again.
If the door is left open for +30 seconds, the alarm will occur.
- If door closed during alarming, alarm will stop immediately.
- Pressing mute button when door open will temporarily delay the alarm for 3 minutes.

6. Service and Contact

Service on the instrument should only be performed by qualified service personnel. To request service
or technical support, please contact Benchmark Scientific or your local Benchmark Scientific
representative.
Ph: +1-908-769-5555
Fax: +1-732-313-7007
Em: Info@BenchmarkScientific.com
Web:
www.BenchmarkScientific.com
P.O. Box 709 Edison, NJ 08817, USA
Page 12 / 12

Hide quick links:

Advertisement

loading
Need help?

Need help?

Do you have a question about the MyTemp Mini CO2 and is the answer not in the manual?

Questions and answers