INSTRUCTION MANUAL
HI759
Maple Syrup Digital Grader
Thank You
Thank you for choosing a Hanna Instruments product. Please read
this instruction manual carefully before using the instrument.
For more information about Hanna and our products,
visit www.hannainst.com.
For technical support, contact your local Hanna office or email
us at tech@hannainst.com
Find your local Hanna office on www.hannainst.com
Preliminary Examination
Please examine this product carefully. Make sure that the instrument is not
damaged. If any damage occurred during shipment, please notify your Dealer.
Each HI759 meter is supplied complete with:
(3) Sample cuvettes and caps
Glycerol standard cuvette
(3) 5 mL syringes with tips
1.5V AAA battery
Instruction manual and quick reference guide
Specifications
Range
golden, amber, dark, very dark
Light Source
light emitting diode @ 560 nm
Light Detector
silicon photocell
Environment
0 to 50°C (32 to 122°F); RH 95% max
Battery Type
1 x 1.5V AAA
Auto-Off
after 10 minutes of non-use
Dimensions
86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5")
Weight
64 g (2.3 oz)
Functional Description
Cover
Cuvette with cap
Cuvette holder
LCD Liquid Crystal Display
Operation button
Errors and Warnings
Out of Range: The sample is out of the measurement
range. Check to make sure you use the sealed glycerol
reference cuvette to calibrate the meter before taking
a measurement.
Inverted Cuvette: The glycerol reference cuvette and
sample have been inserted in the wrong order. Otherwise,
check if the cuvette has been inserted upside down.
Low Battery: The battery is low and must
be replaced soon. Normal operation of the
instrument will not be interrupted.
Dead Battery: The battery is dead and must be
replaced. Once the battery display is flashing, normal
operation of the instrument will be interrupted.
Change the battery and restart the meter.
Light High: There is too much light to
perform a measurement. Please check that
the cover is fully closed, and that the glycerol
reference cuvette was used properly.
Light Low: There is not enough light to perform a
measurement. Check to see if the light is blocked or
dirty. Do not use a syrup sample to calibrate the meter.
Error: Internal error. Please contact your
local Hanna Customer Service Center.

Advertisement

loading
Need help?

Need help?

Do you have a question about the HI759 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Hanna Instruments HI759

  • Page 1 Functional Description Thank You INSTRUCTION MANUAL Cover Thank you for choosing a Hanna Instruments product. Please read HI759 Cuvette with cap this instruction manual carefully before using the instrument. For more information about Hanna and our products, Cuvette holder Maple Syrup Digital Grader visit www.hannainst.com.
  • Page 2 The maple leaf icon will blink to show Measurement that calibration is taking place. HI759 is warranted for a period of one year after date of purchase against defects in workmanship and materials when used for their intended The display will then show “INSERT purpose and maintained according to instructions.