Hanna Instruments HI7630-28 Instruction Manual

Hanna Instruments HI7630-28 Instruction Manual

Conductivity and temperature industrial smart probes

Advertisement

Quick Links

INSTRUCTION MANUAL
HI7630-28
HI7630-48
Conductivity and Temperature
Industrial Smart Probes
HI510 Compatible
®
Hanna
is committed to developing and deploying digital solutions
with a positive impact on the environment and climate.
Please scan the QR code or follow the link
to download the
HI510
https://manuals.hannainst.com/HI510
Hanna Instruments Inc., 584 Park East Drive, Woonsocket, RI 02895 USA
controller user manual.
www.hannainst.com

Advertisement

Table of Contents
loading

Summary of Contents for Hanna Instruments HI7630-28

  • Page 1 Please scan the QR code or follow the link to download the HI510 controller user manual. https://manuals.hannainst.com/HI510 Hanna Instruments Inc., 584 Park East Drive, Woonsocket, RI 02895 USA www.hannainst.com...
  • Page 2: Table Of Contents

    All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA. Hanna Instruments reserves the right to modify the design, construction, or appearance of its products without advance notice.
  • Page 3: Preliminary Examination

    Universal Process Controller. The probes are recommended for a wide range of water applications, from pure and ultrapure water (two-ring series i.e. HI7630-28) to industrial process water (four-ring series i.e. HI7630-48 The two-ring probes can be calibrated using a standard with a value close to the measurement value whereas the four-ring probes provide stable measurements over a wide measure range and do not require frequent calibrations.
  • Page 4: Main Features

    Main Features TDS is a calculated value based on the conductivity of the solution (TDS = factor x EC ). A TDS factor is a conversion factor used to change an EC measurement to a ppm measurement. Sal psu: The practical salinity of seawater relates the ratio of electrical conductivity of a normal seawater sample at 15 ºC and 1 atmosphere to a potassium chloride solution (KCl) with a mass of 32.4356 g/Kg water at the same temperature and pressure.
  • Page 5: Specifications

    5  Specifications 6. SPECIFICATIONS HI7630-28zz Two-Electrode Probe 0.000 µS/cm to 30.00 mS/cm* Range 0.000 mg/L to 15.00 g/L (TDS factor 0.5)* RES  34 Ω cm to 99.99 MΩ • • 0.001, 0.01, 0.1 µS/cm; 0.001, 0.01 mS/cm Resolution** 0.01, 0.1 ppm; 0.001, 0.01, 0.1 g/L RES ...
  • Page 6 Specifications  HI7630-48zz Four-Ring Probe 0.0 µS/cm to 999.9 mS/cm* 0.0 mg/L to 400.0 g/L (TDS factor 0.5)* Range RES  1.00 Ω cm to 9.99 MΩ • • Seawater Salinity 400.0 %NaCl, 42 psu, 80 ppt 0.1 µS/cm; 0.001, 0.01, 0.1 mS/cm Resolution** RES  0.1, 1 Ω cm; 0.01, 0.1, 1 kΩ...
  • Page 7: Probe Dimensions & Cable Connection

    Probe Dimensions & Cable Connection 7. PROBE DIMENSIONS & CABLE CONNECTION Two-Electrode Probes Ø 10 mm (0.4”) 3/4’’ NPT Ø 28.4 mm (1.1”) HI7630-28zz 23 mm (0.9”) 45 mm (1.8”) 189 mm (7.4”) HI7630-2800 189 mm (7.4”) Four-Ring Probes Ø 12 mm (0.5”) 3/4’’...
  • Page 8: Installation

    Installation 8. INSTALLATION General considerations • Probes are easily installed using the 3/4” NPT external thread. • Hand tighten the probe in position. Then, depending on the process, tighten one or two turns with a wrench to secure in place. Do not exceed 10 N m (7.3 lb-ft) torque specification for the probe sensor.
  • Page 9 Installation Installation tee • Install the probe in a tee, horizontally, and direct the flow into the opening at the bottom of the probe. • Upward flow direction must always be ensured. • Maintain a stable flow rate to minimize interferences from bubbles and settling of particulates. •...
  • Page 10 Installation To install the probe: A. Unscrew the protective end cap (9). The cap allows for quick and simple probe maintenance and replacement. B. Slide out the electrode holder cap (3). C. Screw the probe into the internally threaded protective cap (9), step I. D.
  • Page 11: Wiring The Probe To The Controller

    Wiring the Probe to the Controller 9. WIRING THE PROBE TO THE CONTROLLER 1. With the controller disconnected from power, run the probe cable through the conduit opening. 2. Connect the probe leads to the removable terminal connector marked PROBE. Follow the lead markings (+) positive / (–) negative to ensure correct wiring position for the output leads.
  • Page 12: Calibration

    Calibration 10. CALIBRATION HI510 controller allows two types of EC calibration procedures: • Standard two-point conductivity calibration with standards for cell factor determination: 0.000 µS/cm for offset 84.0 µS/cm, 1413 µS/cm, 5.00 mS/cm,12.88 mS/cm for the 0.1/cm cell 80.0 mS/cm, 111.8 mS/cm additional standards for the 1.0/cm cell •...
  • Page 13 25 °C. The self-ionization of water is highly temperature dependent. • Recommended temperature compensation setting for these type of measurements (taken with HI7630-28 probes only) is "Standard" as it utilizes the correct compensation algorithm. Calibration recommendations • Remove the probe from the process and shake all the water from the probe.
  • Page 14: Measurements In Ultrapure Water, Hi7630-28Zz Series Only

    Maintenance 7. Press Process when available. 8. Use directional keys to adjust the process value in agreement with the standard value. 9. Press CFM to confirm and save calibration. Installation recommendations HI510 controller together with a HI7630-28zz probe is designed to meet the ASTM D5391- 99 requirements for electrical conductivity and resistivity of flowing high-purity water samples.
  • Page 15: Accessories

    Accessories 13. ACCESSORIES Ordering Information Product Description Conductivity Solutions HI7030M or HI7030L 12880 µS/cm standard solution, 230 or 500 mL HI7031M or HI7031L 1413 µS/cm standard solution, 230 or 500 mL HI7033M or HI7033L 84 µS/cm standard solution, 230 or 500 mL HI7034M or HI7034L 80000 µS/cm standard solution, 230 or 500 mL HI7035M or HI7035L...
  • Page 16: Abbreviations

    If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization (RGA) number from the Technical Service department and then send it with shipping costs prepaid.

This manual is also suitable for:

Hi7630-48