Dear Customer,
Thank you for choosing a Hanna Instruments Product.
Please read this instruction manual carefully before using the instrument. If you need
HI 755
additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
Preliminary examination:
Marine Alkalinity
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occured during shipment, please notify your Dealer.
Each HI 755 meter is supplied complete with:
• Two Sample Cuvettes and Caps
• 1 bottle of Alkalinity Liquid Reagent
• 1 mL syringe with tip
• 1 x 1.5V AAA Battery
• Instruction Manual
Range
Resolution
Accuracy
Typical EMC Dev.
Light Source
Light Detector
Method
Environment
Battery Type
Auto-Shut off
Dimensions
Weight
For more details about spare parts and accessories
see "Accessories".
Technical specifications:
0 to 300 ppm
1 ppm
±5% of reading ±5 ppm @ 25°C
±1 ppm
Light Emitting Diode @ 610 nm
Silicon Photocell
Colorimetric method. The reaction causes a distinctive
range of colors from yellow to green to greenish blue to
develop.
0 to 50°C (32 to 122°F);
max 95% RH non-condensing
1 x 1.5V AAA
After 10 minutes of non-use.
81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5")
64 g (2.25 oz.).
Functional description:
1. Dust cover.
2. Cuvette with cap.
3. Cuvette holder.
4. Liquid Crystal Display.
5. Button
Errors and warnings:
Light High: There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement.
Please check the preparation of the zero cuvette.
Inverted cuvettes: The sample and the zero cuvette are inverted.
Under range: A blinking "0" indicates that the sample absorbs less
light than the zero reference. Check the procedure and make sure you
use the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery low: The battery must be replaced soon.
Dead battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the
meter.

Advertisement

loading

Summary of Contents for Hanna Instruments HI 755

  • Page 1 Please examine this product carefully. Make sure that the instrument is not damaged. If any damage occured during shipment, please notify your Dealer. 3. Cuvette holder. Each HI 755 meter is supplied complete with: 4. Liquid Crystal Display. • Two Sample Cuvettes and Caps 5.
  • Page 2 • Whenever the cuvette is placed into the measurement cell, it must be dry all segments will be displayed. When the HI 755-26 Liquid Reagents for 25 Alkalinity tests, 1 syringe and 1 tip outside, and completely free of fingerprints, oil or dirt. Wipe it throughly with display shows ‘’Add’’, ‘’C.1’’...