Chino IR-BZ Series Instructions Manual

Compact infrared thermometer
Table of Contents

Advertisement

Quick Links

INST.No.INE-559-P1
IR-BZ Series
COMPACT INFRARED
THERMOMETER
Model: IR-BZP
Always keep this instruction with the unit.
Please be sure to deliver these instructions
with the unit to the end user.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the IR-BZ Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Chino IR-BZ Series

  • Page 1 INST.No.INE-559-P1 IR-BZ Series COMPACT INFRARED THERMOMETER Model: IR-BZP Always keep this instruction with the unit. Please be sure to deliver these instructions with the unit to the end user.
  • Page 3: Table Of Contents

    INST.No.INE-559-P1 Contents PREFACE/ Request and notices Important explanation Warning and attention for the security 1. Introduction ..............1 9.3 Analog dummy output........... 19 1.1 Outline..............1 9.3.1 Analog dummy output1........19 2. Model ................1 9.3.2 Analog dummy output2........20 2.1 Model ..............
  • Page 4 2. If a question has arisen or if an omission was found in this manual, please contact your nearest CHINO’s sales agent or your dealer. 3. Copy right of this instruction manual belongs to CHINO Corporation. All or part of this instruction manual shall not be released to the third party.
  • Page 5 2. Storage ●Do not store the thermometer in hot and humid places. ●For failures of the thermometer, don’t overhaul it by yourself, and contact your sales agent of CHINO Corporation. 3. Disposal ●When you discard this product, please obey the regulation of each local government.
  • Page 6 Don’t use the thermometer if it is broken, smoking or abnormal order is detected. ! These may cause fire. For such abnormal conditions, turn off the power supply immediately and contact your sales agent of CHINO Corporation. ! Disassembling or modifying this thermometer may not only cause failure but also be dangerous to you.
  • Page 7: Introduction

    INST.No.INE-559-P1 1. Introduction 1.1 Outline IR-BZP "Compact Infrared Thermometer" is composed of a Main unit, which CPU is built-in and various function settings are available from digital display and key operation, and Detector. Heat resistance for a Detector is 100°C and it is also water-proof (IP65 equivalency) so it is excellent in environmental resistance.
  • Page 8: Names And Functions Of Component Parts

    INST.No.INE-559-P1 3. Names and functions of component parts 3.1 Main unit 3.1.1 Overview (1) Digital display (6) Hole for installation (2) Function key (3) Event lamp (4) Cable gland Detector side (5) Cable gland Power supply and signal line side (6) Hole for installation Main display: LCD 4-digits, Sub-display: LCD 4-digits, Status marker: Alarm status...
  • Page 9: Function Key And Digital Display

    INST.No.INE-559-P1 3.1.2 Function key and Digital display Function key CONT P EAK (4) Main marker (1) Select key (5) Status marker º C (2) Up key Main display º F ε (7) Sub-marker (3) Down key (6) Measurement (6) Measurement Sub display unit display unit display...
  • Page 10: Installation

    INST.No.INE-559-P1 4. Installation 4.1 Installation of Main unit Fix the Main unit by two installation holes. Caution Do not use the thermometer at following places where; ●Dust, particles and corrosive gas are in the atmosphere ●Electric noise and static electricity are generated frequently ●Ambient temperature is 65°C or higher or -10°C or lower ●Ambient temperature changes widely or humid is high ●Close to high frequency line or power line...
  • Page 11: Measuring Diameter And Measuring Distance

    INST.No.INE-559-P1 4.3 Measuring Diameter and Measuring Distance φ16 φ8 Unit: mm...
  • Page 12: Connections And Wirings

    INST.No.INE-559-P1 5. Connections and Wirings 5.1 Connection to the Main unit ! W a rn i ng To prevent an electric shock, turn OFF the power source for wiring. Detector is connected prior to a shipment. Connect a cable at the power supply and signal line. Use dedicated cable or commercial cable (twisted-pair cable with a shield, outer diameter φ6.5 to φ8.0, wire size 0.2 to 0.5mm2.).
  • Page 13: Connections To Input And Output Terminals

    INST.No.INE-559-P1 5.2 Connections to input and output terminals 5.2.1 Terminal arrangement and terminal nameplate Connecting cable IR-ZBP (Color) (Marker) Grounding Gray 8-36VDC 8-36VDC Power supply 0VDC White/Red 0VDC Black Communication White/Black Contact input Yellow Contact output White/Yellow Common White/Brown A GND Analog ground White/Green A GND...
  • Page 14: Operating

    INST.No.INE-559-P1 6. Operating When installation and connections are completed, turn ON the power supply. 6.1 Digital display CONT PEA K Prepare a measurement by displaying “- - - -” on main display and “lnit” on sub display. Measured value is displayed on the main display when measurement is started.
  • Page 15: Start Option/Parameter Diagram

    INST.No.INE-559-P1 7. Start option/Parameter diagram 7.1 Start option By key operation at start-up, following operation is performed. Chapter of Key operation Display mode Remarks this manual Press SEL key at Parameter setting at Operating mode measurement mode operating mode Long press SEL key at Parameter setting at Engineering mode engineering mode...
  • Page 16: Parameter Diagram

    INST.No.INE-559-P1 7.3 Parameter diagram (Operating mode・Engineering mode) Initial screen Measurement mode Long press on the SEL key Operating Engineering mode mode Time constant Emissivity Key lock Select digit after setting Set/ Release hold function decimal point Transmittance Analog setting Contact input output no, SAMP Select...
  • Page 17: Operating Mode

    INST.No.INE-559-P1 8. Operating mode 1) [Measurement mode screen] is displayed by turning ON the power supply. 2) Press SEL key to display [Operating mode screen] on the right. Operating mode diagram For how to set/select parameters, refer to each chapter. Measurement mode Transition at pressing SEL key is shown on the right.
  • Page 18: Details Of Operating Mode

    INST.No.INE-559-P1 ◇Details of operating mode Press SEL key at measurement mode to operating mode. R e m a r k s It returns to the measurement mode by long pressing (about 2 sec.) on SEL key. Caution It returns to the measurement mode if no key operation is performed for 1 minute. 8.1 Emissivity setting [Emissivity] 1) Press the SEL key from measurement mode screen to display...
  • Page 19: Signal Modulation Mode Selection

    INST.No.INE-559-P1 8.3 Signal modulation mode selection [dELy] Select form to modulate measurement signal (original signal). Select "dELy" to display REAL that is based on original signal CONT PEAK and set modulation degree to 0.0. 1) Press the SEL key couple of times from measurement mode screen to display "Modu"...
  • Page 20: Modulation Degree Setting

    INST.No.INE-559-P1 8.4 Modulation degree setting Caution Modulation degree setting is used with "8.3 Signal modulation selection " If "dELy" is selected in signal modulation mode selection, degree of first order lag (modulation time constant) can be adjusted by setting modulation time constant. If "PEAk"...
  • Page 21: Alarm Setting

    INST.No.INE-559-P1 8.5 Alarm setting Select alarm type (high, low or no alarm) and set temperature alarm point (only one point, high or low alarm). If "ALM" is selected at contact output, contact point is output based on the set value. In other case, AL and AH of status marker lights up only.
  • Page 22: Engineering Mode

    INST.No.INE-559-P1 9. Engineering mode [Measurement mode screen] is displayed by turning ON the power supply. Long press on the SEL key to display [Engineering mode screen] on the below. For how to set/select parameters, refer to each chapter. Transition at pressing SEL key is shown below. *It returns to the measurement mode by long pressing (about 2 sec.) on SEL key.
  • Page 23: Details Of Engineering Mode

    INST.No.INE-559-P1 ◇Details of engineering mode Long press on the SEL key at measurement mode screen to display engineering mode screen. R e m a r k s It returns to the measurement mode by long pressing (about 2 sec.) on SEL key. Caution It returns to the measurement mode if no key operation is performed for 1 min.
  • Page 24: Selection Of Analog Output

    INST.No.INE-559-P1 9.2 Selection of analog output [Analog output types] 1) Press the SEL key couple of times from engineering mode screen 0-20 :0 to 20mA to display "OutS" shown figure below on sub display. 4-20 :4 to 20mA 2) Press ↑ key or ↓ key and change analog output type. tCJ :J thermocouple* tCK :K thermocouple* Selection is completed by changing the display.
  • Page 25: Setting Of Analog Output Maximum Value

    INST.No.INE-559-P1 9.2.2 Setting of analog output maximum value (This is valid only if "0-20" or "4-20" is selected for analog output selection) 1) Press the SEL key couple of times from engineering mode [Analog output maximum value] screen to display "OutH" shown on the right figure on sub CONT PEAK display.
  • Page 26: Analog Dummy Output2

    INST.No.INE-559-P1 9.3.2 Analog dummy output2 (This is valid only if "tCJ", "tCk", "tCr" or "tCS" is selected for analog output selection) 1) Press the SEL key couple of times from engineering mode [Analog dummy output2] screen to display "Out2" shown on the right figure on sub CONT PEAK display.
  • Page 27: Selection Of Hold Function

    INST.No.INE-559-P1 9.4 Selection of hold function Caution Alarm judgment is done by a value that is processed as hold. (Refer to 8.5 Alarm setting) Caution Hold processing is performed to modulating signal. (Refer to 8.3 Signal modulation mode selection) Select whether to use hold function or not. If using the function, select it from sample, peak or valley hold.
  • Page 28: Selection Of Hold Reset Type

    INST.No.INE-559-P1 9.4.1 Selection of hold reset type (Register this function if peak or valley hold is selected at previous section.) [No reset] 1)After register of "PEAK" or "VALL" at the previous section, press the SEL key to display "H.rSt" shown on the right CONT PEAK figure on sub display.
  • Page 29: Setting Of Hold Reset Time

    INST.No.INE-559-P1 9.4.2 Setting of hold reset time Register this function if "in" (internal reset) is selected at previous section. Set hold reset time. [Hold reset time] CONT PEAK 1) Press the SEL key couple of times from engineering mode screen to display "r.tiM" shown on the right figure on sub display.
  • Page 30: Setting Of Time Constant Digit After

    INST.No.INE-559-P1 9.6 Setting of time constant digit after decimal point 1) Press the SEL key couple of times from engineering mode [Time constant digit after decimal point] screen to display "tAu.d" shown on the right figure on sub CONT PEAK display.
  • Page 31: Selection Of Contact Input

    INST.No.INE-559-P1 9.7 Selection of contact input [No contact input] Select a function to use contact input. CONT PEAK 1) Press the SEL key couple of times from engineering mode screen to display "din" shown on the right figure on sub display.
  • Page 32: Selection Of Contact Output

    INST.No.INE-559-P1 9.8 Selection of contact output [No contact output] Select a signal to output at contact output. CONT PEAK 1) Press the SEL key couple of times from engineering mode screen to display "do1" shown on the right figure on sub display.
  • Page 33: Setting Of Device Address

    INST.No.INE-559-P1 9.10 Setting of device address [Device address] 1) Press the SEL key couple of times from engineering mode CONT PEAK screen to display "Adr" shown on the right figure on sub display. 2) Press ↑ key or ↓ key and change set value of device address. Setting is completed by changing the numeric value.
  • Page 34: Self-Diagnostic Function And Overflow/Underflow Indication

    INST.No.INE-559-P1 10. Self-diagnostic function and Overflow/underflow indication 10.1 Self-diagnostic function Self-diagnostic function is built-in to this thermometer and an error number is displayed at abnormal conditions. Output Sub display Contents Countermeasures (Note) Abnormal Detector Check if the Detector is overheated or ○...
  • Page 35: Maintenance

    INST.No.INE-559-P1 11. Maintenance 11.1 Regular inspection Perform following check-ups regularly or depending on the situation. 11.2 Troubleshooting If a trouble is recognized, check following items and perform troubleshooting. 11.2.1 Measuring value not displayed or displayed lower Inspection items Countermeasures 1) Check that power is supplied to the thermometer Supply normal power supply voltage.
  • Page 36: Specifications

    INST.No.INE-559-P1 12. Specifications 12.1 Specifications □ Model IR-BZPHGE Measuring system Wide-band infrared radiation thermometer Detecting element Thermopile 8 to 14μm Measuring wavelength Measuring range 0 to 1000°C Below 20°C : ±2°C Accuracy ratings *1 20°C or higher , below100°C : ±1°C 100°C or higher : ±1% of measured value Reproducibility...
  • Page 37: Parameter Selection/ Setting

    INST.No.INE-559-P1 13. Parameter selection/ setting 13.1 List of parameter selection/ setting (operating mode) Selection/ setting Chapter of Setting range Default setting item display this manual ε Emissivity setting 1.999 to 0.100 1.000 Transmittance 1.999 to 0.100 1.000 setting Signal modulation Modu dELy, PEAk, VALL dELy...
  • Page 38: List Of Parameter Selection/ Setting

    INST.No.INE-559-P1 13.2 List of parameter selection/ setting (engineering mode) Chapter of Selection/ setting Setting range Default setting this item display manual Key lock setting/ LCK.S YES, no release 0-20 : 0 to 20mA 4-20 : 4 to 20mA Analog output : J thermocouple OutS 4-20mA...
  • Page 39: Initialization Method

    INST.No.INE-559-P1 14. Initialization method 1) On measurement mode screen, press and hold ↓ key+SEL key for 2 sec. 2)"init" in main display blinks. 3) Press and hold for 3sec. 4) Blinking " init " lights up and initialization is completed. *If keys are released while "init"...
  • Page 40: External Dimensions

    INST.No.INE-559-P1 15.External dimensions 15.1 IR-BZP Main unit (12) (17) Cable gland Power supply, signal cable inlet Compatible cable dia.φ6.5 to φ8 Cable gland Mounting hole Detector cable inlet 2-φ4.5 2-M3×4DP Unit:mm Mounting screw for DIN rail mounting base(IR-ZBDP) -34-...
  • Page 41: Ir-Bzp Detector

    INST.No.INE-559-P1 15.2 IR-BZP Detector Detector cable (1.5, 3, 8 or 15m) OD :φ3.3 Heat resistant : 100℃ M12 P1 Unit:mm 15.3 Connecting cable IR-ZBP□□□ □□□ 002:2m O chip M3.5 005:5m Shield cable OD :φ6.6 Heat resistant : 80℃ Select 2m or 5m Color Marker Black...
  • Page 42: Mounting Bracket For Detector Ir-Zbmp

    INST.No.INE-559-P1 15.4 Mounting bracket for Detector IR-ZBMP 90° φ3.5 Unit:mm 15.5 DIN-rail mounting base for Main unit IR-ZBDP 2-φ4 6-M3 countersunk (not used) Mounting hole for Main unit Unit:mm -36-...
  • Page 44 32-8, KUMANO-CHO, ITABASHI-KU, TOKYO 173-8632 Telephone: +81-3-3956-2171 Facsimile: +81-3-3956-0915 Web site http://www.chino.co.jp/ INE-559-P0 Feb-2017 IR-BZSeries COMPACT INFRARED THERMOMETER Model : IR-BZP Printed in Japan...