Chino IR-CA series Instructions Manual

High function type infrared radiation thermometer
Hide thumbs Also See for IR-CA series:
Table of Contents

Advertisement

Quick Links

INST.No.INE-454-P2
IR-CAQ Series
HIGH FUNCTION TYPE INFRARED
RADIATION THERMOMETER
Model: IR-CAQ
Always keep these instruction with the unit.
P l e a s e b e s u r e t o d e l i v e r t h e s e i n s t r u c t i o n s
w i t h t h e u n i t t o t h e e n d u s e r.
Software Ver.1.2
[Software volume]

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the IR-CA series and is the answer not in the manual?

Questions and answers

Summary of Contents for Chino IR-CA series

  • Page 1 INST.No.INE-454-P2 Software Ver.1.2 IR-CAQ Series HIGH FUNCTION TYPE INFRARED RADIATION THERMOMETER [Software volume] Model: IR-CAQ Always keep these instruction with the unit. P l e a s e b e s u r e t o d e l i v e r t h e s e i n s t r u c t i o n s w i t h t h e u n i t t o t h e e n d u s e r.
  • Page 2 CHINO Corporation. 3. CHINO Corporation shall not be liable for any operation results. Preface To use high function type infrared radiation thermometer...
  • Page 3: Table Of Contents

    3. Details of the operator mode··············································· 3 4.9.3-1) Zero/span adjustments execution selection ················· 17 3.1 Emissivity (ratio) programming ······································· 3 4.9.3-2) Zero side data display/setting by IR-CA series············ 18 3.1.1 Emissivity (ratio) programming 1································· 3 4.9.3-3) Zero side data display/setting by furnace 3.1.2 Emissivity (ratio) programming 2·································...
  • Page 4: Operation

    INST.No.INE-454-P2 Software Ver.1.2 1. Operation 1.1 Digital display After the installation, the connections and the CONT PEAK wirings are completed, turn all mains power on. Measurement starts and the measured value is displayed in the main display. When the low alarm or the high alarm activates, the status marker “AL”...
  • Page 5: Parameter Programming/Selection (Engineering Mode)

    INST.No.INE-454-P2 Software Ver.1.2 2. Operating mode and engineering mode 2.2 Parameter programming/selection (Engineering mode) [ Measurement mode ] In the measurement mode, press SEL key and ENT key simultaneously to display the engineering mode shown CONT PEAK in the right figure for programming or selecting a parameter.
  • Page 6: Operating Mode

    INST.No.INE-454-P2 Software Ver.1.2 3. Operating mode 3. Details of the operator mode 3.1 Emissivity(ratio) programming If the emissivity (ratio) of object measured is low, the temperature displayed becomes lower than the exact temperature and the emissivity (ratio) is to be compensated. But programming differs by the selected measurement wavelength.
  • Page 7: Emissivity (Ratio) Programming 2

    INST.No.INE-454-P2 Software Ver.1.2 3. Operating mode 3.1.2 Emissivity (ratio) programming 2 : Program the emissivity (ratio) of 1.35/0.9µm in “2col” and 0.9µm in “1col” in case of automatic emissivity calculation. In case of automatic emissivity calculation, program the emissivity ratio at high temperature side range. Even if it sets up in case of 2 color measurement and single color measurement it is not related to measurement.
  • Page 8: Automatic Emissivity Calculation

    INST.No.INE-454-P2 Software Ver.1.2 3. Operating mode 3.2.2 Automatic emissivity calculation (External …Model IR- CAQ 5 only) : Selection of external programming In the measurement mode, the emissivity is automatically calculated by entering the reference temperature value by the analog input (option). Reference Necessary to program the scaling of the analog input in [4.11 Analog input programming])
  • Page 9: Modulation Degree Programming

    INST.No.INE-454-P2 Software Ver.1.2 3. Operating mode 3.4 Modulation degree programming Caution Execute the programming of the modulation degree together with [3.3 Signal modulation mode selection]. When the "dELy" is selected in the signal modulation mode, the first-order lag degree can be adjusted.
  • Page 10: Alarm Programming

    INST.No.INE-454-P2 Software Ver.1.2 3. Operating mode 3.5 Alarm programming (Model IR-CAQ J only contact output possible: As for other models “AL” , “AH” of the status marker are lighted.) This programming is for the selection of the alarm type (high alarm, low alarm or alarm disable) and the programming an alarm set-point (1 point only in the high alarm or the low alarm).
  • Page 11: Engineering Mode

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode 4. Engineering mode In the measurement mode, press SEL and ENT keys simultaneously to display the engineering mode screen. Refer to the engineering mode screen in [6.3 Parameter programming/selection (Engineering mode)] Caution To return to the measurement mode screen from the engineering mode screen •...
  • Page 12: Analog Dummy Output Programming

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode 4.2 Analog dummy output programming This programming is for analog dummy output based on programmed ratio. The output of 0 to 100% is corresponding to 4 to 20mA. Note that this screen will not move to the measurement mode even if any key is not pushed for 1 minute.
  • Page 13: Hold Functions (Hold Disable, Sample Hold Peak Hold) Selection

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode 4.3 Hold functions (Hold disable, sample hold, peak hold) selection This mode is for the selection of the hold function (hold function disable, sample hold or peak hold). When the peak hold is selected, [4.3.2 Peak hold reset type (time reset or remote contact reset) selection] is necessary.
  • Page 14: Peak Hold Reset Type Selection

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode 4.3.2 Peak hold reset type selection (Selection of the outside reset・・・Model IR-CAQ J only) (It is necessary to select when the peak hold is selected in [4.3.1 Hold function type selection].) 1) After the selection of “PEAk” in [4.3.1 Hold function type [Reset disable] selection], press SEL key until “HrSt”...
  • Page 15: Peak Hold Reset Time Programming

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode 4.3.3 Peak hold reset time programming (It is necessary to program the reset time when the peak hold is selected in [4.3.1 Hold function type selection] and then “in” (internal reset) is selected in [4.3.2 Peak hold reset type selection].) 1) After the selection of “in”...
  • Page 16: Modulation Time Constant Unit Selection

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode 4.5 Modulation time constant unit selection [3.4.1 Modulation time constant programming] to the cipher below the decimal point of when that sets up the fixed number is selected when it puts it. 1) To select the measurement unit, press SEL key in the CONT PEAK engineering mode screen until “tau.d”...
  • Page 17: Automatic Emissivity Calculation Type Selection (External Setting: Model Ir-Caq 5 Only)

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode 4.7 Automatic emissivity calculation type selection (External setting :Model IR-CAQ 5 only) This mode is for the selection of the automatic emissivity calculation function (disable or enable). In case of using the automatic emissivity calculation functions, select the internal or external setting. 1) To select the automatic emissivity calculation type, press [Calculation disable] SEL key in the engineering mode screen until “E.Aut”...
  • Page 18: Measuring Wavelength Selection

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode 4.8 Measuring wavelength selection This mode is the function that selects measurement wavelength. 1) To enter the Wavelength selection, press SEL key in the CONT PEAK engineering mode screen until “dEtc” appears in the sub display as shown in the right figure.
  • Page 19: Zero/Span Adjustments

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode 4.9 Zero/span adjustments This mode is for the zero and span adjustments of your thermometer by entering the black body temperature at the minimum temperature range (zero side) and the maximum temperature range (span side) by using your block body furnace. 4.9.0 Procedure for Zero/span adjustments 1) It is necessary to measure the furnace temperature at the zero side and the span side by a reference radiation thermometer in advance.
  • Page 20: Zero/Span Adjustments (2)

    4.9.3 Correction after the zero/span adjustments. (measured value became abnormal) The IR-CA series measured value at the time of zero/span adjustments and also the furnace temperature at the zero side/span side by a reference radiation thermometer [4.9.0 Procedure for Zero/span adjustments] are able to confirm and correct.
  • Page 21: 3-2) Zero Side Data Display/Setting By Ir-Ca Series

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode 4.9.3-2) Zero side data display/setting by IR-CA series 1) When entering into [Zero side data display/setting by IR-CA CONT PEAK series], “MES.Z” appears in the sub display as shown in the right figure. Press SEL key in the case of only the confirmation of data and move to [4.9.3-3) Zero side data display/setting by furnace...
  • Page 22: 3-4) Span Side Data Display/Setting By Ir-Ca Series

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode 4.9.3-4) Span side data display/setting by IR-CA series radiation thermometer 1) To enter the span side data display/setting by IR-CA series [Span side] radiation thermometer, press SEL key in the [4.9.3-3) Zero side CONT...
  • Page 23: Selection Of Correction After The Zero/Span

    1) To enter the Selection of for correction after the zero/span [Modifying programmed adjustments programming item, press SEL key in the [4.9.3-4) desired value] Span side data display/setting by IR-CA series] screen. CONT PEAK “End” appears in the sub display as shown in the right figure.
  • Page 24: Correction Programming

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode 4.10 Temperature display/Analog output correction programming This mode is for the correction of temperature display and analog output by knee-points (up to 8 points). It is necessary to program from 4.10.1 to 4.10.5 continuously. 4.10.1 Programming of number of temperature display and analog output correction data (number of knee-point) Program the number of the knee-point.
  • Page 25: Analog Output Correction

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode 4.10.3 Entering of data before temperature display and analog output correction Enter the data before the temperature display and the analog output correction. 1) After the programming of the temperature display and the [Entering of data before analog output correction data No.
  • Page 26: Selection Of Temperature Display And Analog Output Correction For Programming Item

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode 4.10.5 Selection of temperature display and analog output correction for programming item This mode is for the selection of “Modifying programmed desired value”, “Canceling programmed desired value and exiting from this mode” or “Execution of correction command” of above [4.10.1] up [4.10.4] in the [4.10 Temperature display/Analog output correction programming] 1) To enter the Selection of temperature display and analog output [Modifying programmed...
  • Page 27: Analog Input Programming (Model Ir-Caq 5 Only)

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode (Model IR-CAQ 4.11 Analog input programming 5 only) 4.11.1 Selection of analog input items This mode is for the selection of the analog input (analog input disable, for remote emissivity programming or for automatic emissivity calculation). [Analog input disable] 1) To select the analog input item, press SEL key in the CONT...
  • Page 28: Emissivity Scaling For Remote Emissivity Programming

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode 4.11.2 Emissivity scaling for remote emissivity programming This mode is for the scaling of the emissivity value when the remote emissivity programming is selected. [ Minimum value of emisivity 1) To program the minimum emissivity value for the scaling, scaling] press SEL key in the engineering mode screen until “AinL”...
  • Page 29: Temperature Scaling For Automatic Emissivity Calcuiation

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode 4.11.3 Temperature scaling for automatic emissivity calculation This mode is for the scaling of the temperature value when the automatic emissivity calculation is selected. 1) To program the minimum temperature value for the scaling, [Minimum value of press SEL key in the engineering mode screen until “AinL”...
  • Page 30: Communications Programming

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode (Model IR-CAQ 4.12 Communications programming S only) This mode is for the programming of the address (effective in multi-drop connections) of this thermometer and communications speed. Address ] 4.12.1 Address programming CONT PEAK 1) To program the address of this thermometer, press SEL key in the engineering mode screen until “Adr”...
  • Page 31: Communication Protocol Mode Selection

    INST.No.INE-454-P2 Software Ver.1.2 4. Engineering mode 4.12.3 Communication protocol mode selection Select the communication protocol, MODBUS protocol mode or IR-CA series exclusive protocol mode. [MODBUS RTU mode ] 1) To select the communications protocol mode, press SEL key CONT PEAK in the engineering mode screen until “Prot”...
  • Page 32: Self-Diagnostic Function And Overflow/Underflow

    INST.No.INE-454-P2 Software Ver.1.2 Overflow/underflow Self-diagnostic function and 5.1 Self-diagnostic function This self-diagnosis function is built-in. An error No will be displayed on abnormal conditions. Output Contents Countermeasure display (Note*) Abnormal ambient Use the thermometer in the temperature environment from 0 to 50ºC. 2...
  • Page 33: List Of Starting Up Modes

    INST.No.INE-454-P2 Software Ver.1.2 6. List of starting up modes 6.1 Starting up modes Keys Screen Remarks Paragraph Temperature measurement Measurement (Power on) and parameter programming mode in the operator mode Press SEL and ENT keys simultaneously Engineering Parameter programming in in measurement mode screen.
  • Page 34: Parameter Programming/Selection (Engineering Mode)

    Not execution 4.9.3-1) zero/span yES: Execution adjustments. (measured value became abnormal) Zero side data MES.Z 4.9.3-2) display/setting by IR-CA series Zero side data SEt.Z 4.9.3-3) display/setting by furnace temperature by a reference radiation thermometer Span side data MES.S 4.9.3-4)
  • Page 35 INST.No.INE-454-P2 Software Ver.1.2 7. Parameter programming/selection Parameter item Sub display Parameter Default Chapter Zero side data SEt.S 4.9.3-5) display/setting by furnace temperature by a reference radiation thermometer Selection of 4.9.4 cont: Confirm and modify programmed correction after the desired value again. zero/span CAnc: Canceling programmed desired adjustments...
  • Page 36 32-8, KUMANO-CHO, ITABAS HI-KU, TOKYO 173-8632 Telephone: +81-3-3956-2171 Facsimile: +81-3-3956-0915 Web site http://www.chino.co.jp/ INE-454-P2 Jan-'05 IR-CAQ Series High function type infrared radiation thermometer IR-CAQ Printed in Japan...