Kipp & Zonen CVF4 Instruction Manual

Hide thumbs Also See for CVF4:

Advertisement

Quick Links

Instruction Manual
CVF4 Ventilation Unit
V2007-01

Advertisement

Table of Contents
loading

Summary of Contents for Kipp & Zonen CVF4

  • Page 1 Instruction Manual CVF4 Ventilation Unit V2007-01...
  • Page 2: Important User Information

    Dear customer, thank you for purchasing a Kipp & Zonen instrument. It is essential that you read this manual completely for a full understanding of the proper and safe installation, use, maintenance and operation of your new CVF4 Ventilation Unit. We understand that no instruction manual is perfect, so should you have any comments regarding this manual we will be pleased to receive them at: Kipp &...
  • Page 3: Table Of Contents

    5.2. CMF4 Mounting Fixture ........................11 5.3. CMB1 Mounting Bracket ........................11 5.4. Adapter for use of the CVF4 in the CM121C shadow ring ..............12 Maintenance ..............................14 Technical specifications ..........................15 7.1. Optical and electrical ......................... 15 7.2.
  • Page 4: Introduction

    The CVF4 is meant for all CM, CMP, SMP, CG, CGR and CUV radiometers. However with the CMP3 and SMP3 and CGR3 the ventilation effect will be less due to the larger opening around the dome. The CVF4 is meant to run continuously.
  • Page 5: Waste Disposal

    2.2. Waste disposal The pictogram showing a waste bin with a cross means that the product is subject to European Union regulations covering segregated waste disposal. This applies both to the product itself and to any accessories marked with the same symbol. Disposal of any such item as unsorted domestic waste is not allowed.
  • Page 6: Cvf4 Main Components

    3. CVF4 main components In this section the main CVF4 components are described. 3.1. CVF4 top side Air duct and fan Heating elements Mounting holes CMP3, SMP3 and CGR3 Mounting holes all other instruments Connector Cover nuts Slit for radiometer cable 3.2.
  • Page 7: Installation

    Tools required If the CVF4 is mounted with the supplied screws only the supplied Allen key is required. Make sure there is space to fasten the screw under the bottom of the plate. To fasten the radiometers other Allen keys or screwdrivers may be necessary.
  • Page 8: Installation Procedure

    4.4. Installation procedure First the CVF4 is screwed flat on its mounting plate. In case the CVF4 is mounted on a solar tracker, the tracker has to be levelled first. On our sun trackers the CVF4 is positioned such that the fan sticks out over the edge of the plate. This ensures the air inlet is never blocked (see picture of the CVF4 on the SOLYS2 Sun Tracker on the next page).
  • Page 9 (not secured yet). In case the CVF4 is mounted on a solar tracker, finish the alignment of the tracker first before securing the radiometer screws because tracker (and radiometer) levelling might need re-adjustment.
  • Page 10: Electrical Connections

    Shield Earth The ground is common for the ventilator and the heater (red, blue and brown). The CVF4 has an optional 8- wire Kipp & Zonen cable for connecting the 12 VDC to the ventilator and heater. The tacho output provides a 5 Volt square wave signal. It can be connected to a data logger to (remotely) check proper operation of the CVF4.
  • Page 11: Accessories

    5. Accessories 5.1. Spare filters Spare CVF4 filters are available under part number 2682047 from Kipp & Zonen or under EBM-Papst part number: FP80T from a local distributor of Pabst fans. 5.2. CMF4 Mounting Fixture Article number 0362703 CMF4 5.3.
  • Page 12: Adapter For Use Of The Cvf4 In The Cm121C Shadow Ring

    5.4. Adapter for use of the CVF4 in the CM121C shadow ring CVF4 Remove CVF4s feet and Remount two feet in these counterbore holes Relocation of 2 of the CVF4s mounting feet to fit it on the CM121C shadow ring...
  • Page 13 The blue adapter part is included with the CM121C. Note: CM121B (Article number 0346900) is not supplied with, nor aligned for the CVF4 adapter (Article number 0346700) CM121C (Article number 0346901) is aligned for and supplied with the CVF4 adapter...
  • Page 14: Maintenance

    6. Maintenance Regular inspection of the CVF4 is advised. Depending on the location (air pollution) this can be monthly or yearly. When on site for inspection of other instruments it is advised to check the CVF4 filter. At the same time radiometer levelling and desiccant can be checked. For access to the desiccant the 2 mounting nuts have to be loosened and the top cover taken off.
  • Page 15: Technical Specifications

    7. Technical specifications 7.1. Electrical Specifications Heater voltage 12 ± 1 VDC Heater current 460mA for the 7W version Heater power 5.5W Ventilator voltage 12 ± 1 VDC Ventilator current 520mA ±20% Ventilator power 6.3W Tacho signal 5V and 475 Hz ±25% Dimensions 210x178x126 mm Weight packed...
  • Page 16: Trouble Shooting

    8. Trouble shooting Malfunction Cause Check Ventilator and heater do not Power failure Check 12 VDC and/or mains voltage work Ventilator does not work Mechanical obstruction Clean filter and/or remove obstruction Snow below air intake Frozen filter Ventilator does not work Electrical failure Check fan impedance ~30 Ω...