Kipp & Zonen CVF4 Instruction Manual

Ventilation unit
Hide thumbs Also See for CVF4:

Advertisement

Quick Links

Instruction Manual
CVF4 Ventilation Unit
Revision March 2014

Advertisement

Table of Contents
loading

Summary of Contents for Kipp & Zonen CVF4

  • Page 1 Instruction Manual CVF4 Ventilation Unit Revision March 2014...
  • Page 2 Instruction Manual - CVF4...
  • Page 3 WARRANTY AND ASSISTANCE This equipment is warranted by CAMPBELL SCIENTIFIC (CANADA) CORP. ("CSC") to be free from defects in materials and workmanship under normal use and service for ***** twenty-four (24) months from date of shipment unless specified otherwise. ***** Batteries are not warranted.
  • Page 4: Important User Information

    Dear customer, thank you for purchasing a Kipp & Zonen product. It is essential that you read this manual completely for a full understanding of the proper and safe installation, use, maintenance and operation of your new CVF4 Ventilation Unit.
  • Page 5 Instruction Manual - CVF4...
  • Page 6: Declaration Of Conformity

    EN 61326-1:2000 Safety EN 61010-1:2001 Radio part following the provisions of the directives (if applicable): EMC-directive 2004/108/EC Electrical safety 2005/65/EC Following the provisions of the directive. Delft, 1 March 2014 B.A.H. Dieterink President Kipp & Zonen B.V. Instruction Manual - CVF4...
  • Page 7 Instruction Manual - CVF4...
  • Page 8: Table Of Contents

    3.2.1 CVP 1 ................................................3.2.2 CVP 2 ................................................CMF4 Mounting Fixture ........................................ CMB 1 Mounting Bracket ......................................Adapter for use of the CVF4 in the CM 121C shadow ring ........................Maintenance ............................................ Specification ........................................... Optical and electrical ........................................Dimensions and weight ........................................
  • Page 9 Instruction Manual - CVF4...
  • Page 10: Introduction

    - Reduce the zero offset of the radiometer. The CVF4 is meant for all CM, CMP, SMP, CG, CGR and CUV radiometers. However with the CMP 3 and SMP3 and CGR 3 the ventilation effect will be less due to the larger opening around the dome. The CVF4 is meant to run continuously. The heater can either be switched on permanently for cold regions or be switched by a data logger or clock to remove dew in the morning and be switched off...
  • Page 11: Cvf4 Top Side

    1.1 CVF4 top side Ventilator Heater Mounting holes CMP 3, SMP3 and CGR 3 Mounting holes all other instruments Connector Cover nut Slit for radiometer cable 1.2 CVF4 bottom side Filter cover Connector box Mounting feet for SOLYS 2 / Feet for stabilisation...
  • Page 12: Installation

    2.2 Tools required If the CVF4 is mounted with the supplied screws only the supplied Allan key is required. Make sure there is space to fasten the screw under the bottom of the plate. To fasten the radiometers other Allan keys or screwdrivers may be necessary.
  • Page 13: Installation Procedure

    2.4 Installation procedure First the CVF4 is screwed flat on its mounting plate. In case the CVF4 is mounted on a solar tracker, the tracker has to be levelled first. On our sun trackers the CVF4 is positioned such that the fan sticks out over the edge of the plate. This ensures the air inlet is never blocked (see picture of the CVF4 on the SOLYS 2 Sun Tracker on the next page).
  • Page 14 Then the radiometer is mounted in the CVF4 and levelled with its own levelling feet. Afterwards the radiometer mounting screws and shoulder washers are fitted (not secured yet). In case the CVF4 is mounted on a solar tracker, finish the alignment of the tracker first before securing the radiometer screws because tracker (and radiometer) levelling might need re-adjustment.
  • Page 15: Electrical Connections

    Reliez à la terre si le radiomètre n'est pas connecté Conectar a tierra si el radiómetro no lo está The CVF4 has a 4 wire cable for connecting the 12 VDC to the ventilator and heater. The ground is common for the ventilator, heater and the tacho output.
  • Page 16: Accessories

    3. Accessories 3.1 Spare CVF4 filters Article number 2682047 Set of 5 spare ventilation fan inlet filters. 3.2 Power supply Two optional 12 VDC power supplies are available for outdoor (CVP 1) and indoor (CVP 2) use. 3.2.1 CVP 1 Article number 0357700 The CVP 1 is a waterproof (IP 65) power supply for outdoor use with 12 VDC @ 3.5 A output.
  • Page 17: Cmf4 Mounting Fixture

    CMF4 The CMF4 consists of a mounting plate plus rod. It can be used for one CVF4, looking up or 2 CVF4’s where one is looking down. This allows for highly accurate albedo or net-radiation measurements with 2 ventilated radiometers.
  • Page 18: Adapter For Use Of The Cvf4 In The Cm 121C Shadow Ring

    The blue adapter part is included with the CM 121C. CM 121B (Article number 0346900) is not supplied with, nor aligned for the CVF4 adapter (Article number 0346700) CM 121C (Article number 0346901) is aligned for and supplied with the CVF4 adapter...
  • Page 19 Instruction Manual - CVF4...
  • Page 20: Maintenance

    4. Maintenance Regular inspection of the CVF4 is advised. Depending on the location (air pollution) this can be monthly or yearly. When on site for inspection of other instruments it is advised to check the CVF4 filter. At the same time radiometer levelling and desiccant can be checked. For access to the desiccant the 2 mounting nuts have to be loosened and the top cover taken off.
  • Page 21 Instruction Manual - CVF4...
  • Page 22: Specification

    Lower power consumption for both ventilator and heater with improved performance Wide operating voltage range High quality cable with very low voltage drop Note: The performance specifications quoted are worst-case and/or maximum values 5.2 Dimensions and weight 355 mm 230 mm Weight without cable: 1.6 kg Instruction Manual - CVF4...
  • Page 23 Instruction Manual - CVF4...
  • Page 24: Trouble Shooting

    Ventilator does not work Mechanical obstruction Clean filter and/or remove obstruction Ventilator does not work Electrical failure Check fan impedance ~30 Ω Heater does not work Resistor defective Measure power resistors: Specifications: 15 Watt and 50 / 39 Ω Instruction Manual - CVF4...
  • Page 25 Instruction Manual - CVF4...
  • Page 26: Customer Support

    • Interfaces and power supplies • History of any previous repairs or modifications • Pictures of the installation • Overview of the local environment conditions Kipp & Zonen guarantees that your information will not be shared with other organisations. Instruction Manual - CVF4...
  • Page 27 Our customer support remains at your disposal for any maintenance or repair, calibration, supplies and spares. Für Servicearbeiten und Kalibrierung, Verbrauchsmaterial und Ersatzteile steht Ihnen unsere Customer Support Abteilung zur Verfügung. Notre service 'Support Clientèle' reste à votre entière disposition pour tout problème de maintenance, réparation ou d'étalonnage ainsi que pour les accessoires et pièces de rechange.