Philips Respironics E30 User Manual page 39

Ventilation solution
Table of Contents

Advertisement

Alarm
Loss of power
Priority
Device action
Device
Shuts down
inoperative
Power
Shuts down
failure
High
Operates
High
Operates;
if the alarm
continues for
10 seconds, the
alarm escalates
to a device
inoperable
alarm.
Philips Respironics E30 User Manual
User action
Check tube is not crushed or folded
such that air flow is restricted. Check
mask is attached properly and without
any obstruction. If the alarm continues,
contact Philips Respironics or the
product support team.
Immediately connect the ventilator
to an alternate power source. If no
alternate power source is available,
immediately place the patient on an
alternate source of ventilation. Then
check your power connections. Make
sure there is power at the outlet or
power source. If the alarm continues,
contact Philips Respironics or the
product support team.
This could be caused by an excessive
leak, blockage, or a device malfunction.
Press the control dial to silence the
alarm. Press the control dial again to
acknowledge and remove the alarm
from the screen. Remove your mask.
Check for the following: dirty inlet
filters, blocked air intake, excessive
leak in the patient circuit. If the alarm
continues, contact Philips Respironics
or the product support team.
This may be caused by a
malfunctioning device. Press the
control dial to silence the alarm. Press
the control dial again to acknowledge
and remove the alarm from the screen.
Remove mask, and remove power from
the device. Restore power. If the alarm
continues, contact Philips Respironics
or the product support team.
Alarms, alerts, and troubleshooting
33