  • Page 2: Table Of Contents

    CONTENTS 1. Introduction ....................2 2. Specifications ..................2 - 4 3. Symbol Definition and Button Location ..........5 - 6 4. Operation Instructions ..............7 - 9 4.1 Power-Up ....................7 4.2 Connecting the Thermocouples ............7 4.3 Selecting the Temperature Scale ............
  • Page 3: Introduction

    1. INTRODUCTION/ 2. SPECIFICATIONS INTRODUCTION This instrument is a digital thermometer for use with platinum-type temperature sensor. Temperature indication follows IEC751 temperature table for PT-type sensor. SPECIFICATIONS NUMERICAL DISPLAY: 5 digital liquid crystal displays MEASUREMENT RANGE: -100°C ~ 400°C; -148°F ~ 752°F RESOLUTION: 0.01°C;...
  • Page 4: Specifications

    2. SPECIFICATIONS ACCURACY: AT 23 ± 5°C Range Accuracy -100°C ~ 400°C ±(0.05% reading + 0.1°C) -148°F ~ 752°F ±(0.05% reading + 0.2°F) TEMPERATURE COEFFICIENT: For ambient temperatures from 0°C ~ 18°C and 28°C ~ 50°C, for each °C ambient below 18°C or above 28°C add the following tolerance into the accuracy spec.
  • Page 5 2. SPECIFICATIONS BATTERY LIFE: Approx. 100hours STANDARD ACCESSORY: Pt-100 Probe(class A), Instruction manual, 9V Battery, windows software, carrying case, USB cable. OPTIONAL ACCESSORIES: AC adapter, RS-232 Cable.
  • Page 6: Symbol Definition And Button Location

    3. SYMBOL DEFINITION AND BUTTON LOCATION : The Battery is not su cient for proper operation : The Minimum value is now being displayed : The Maximum value is now being displayed : This indicates that the display data is being hold HOLD : This indicates Auto Power O is enabled : Readings are being logged...
  • Page 7 3. SYMBOL DEFINITION AND BUTTON LOCATION Pt type temperature sensor HOLD button connector °C , °F control button LCD display PC Interface Power ON/OFF and Back External DC 9V power supply light button jack Relative readout button Tilt stand REC button Battery Compartment MAX MIN control button...
  • Page 8: Operation Instructions

    4. OPERATION INSTRUCTIONS 4.1 POWER UP Press the “ ” key to turn ON the thermometer and Press and hold “ ” buttons for 3 seconds to OFF. 4.2 CONNECTING THE THERMOCOUPLES For measurement, plug the temperature probe into the input connectors.
  • Page 9: Back Light Operation

    4. OPERATION INSTRUCTIONS 4.5 BACK LIGHT OPERATION Press the button will turn back light on and Press it once again will turn o . The meter will turn back light o if there is no push button for 30 seconds. 4.6 RELATIVE OPERATION When one press the button, the meter will memorize the present...
  • Page 10: Auto Power O

    4. OPERATION INSTRUCTIONS Press button again, MAX and MIN will blink together. This means that all these data is updated in the memory and the reading is the present temperature. One may press to circulate the display mode among these options.
  • Page 11: Temperature Measurement

    5. TEMPERATURE MEASUREMENT 5.1 CORRECT MEASUREMENT METHOD The temperature sensor is located at the end of the metal sheath of the sheath type TEMPERATURE PROBE. To accurately test internal temperature insert the probe into the item you want to measure to a distance at least 15 times the diameter of the sheath 5.2 CONNECTOR CONFIGURATION P.10...
  • Page 12: Setting The Time

    6.SETTING THE TIME Press and hold button and then power on the meter to enter setup mode. Press button to enter Time setting mode. The right two digits blink. ▲or ▼ button until the display shows the correct year, Press and then press button to select.
  • Page 13: Changing The Logging Interval

    7. CHANGING THE LOGGING INTERVAL The logging interval determines how often the thermometer stores logged readings in memory. Press and hold button and then power on the meter to enter setup mode. Press button to enter Interval setting mode. The left two digits blink.
  • Page 14: Clearing Memory

    8. CLEARING MEMORY 9. TEMPERATURE METER CALIBRATION SETUP CLEARING MEMORY When memory is full, REC appears and blinks on the display and logging stops. You can clear memory in next power on the meter to enter clear memory mode. Press and hold button and then power on the meter to enter clear memory mode to delete logged readings from memory.
  • Page 15 9. TEMPERATURE METER CALIBRATION SETUP Press and hold buttons and then power on the meter to enter setup mode. Press button to enter calibration mode.(in 3 seconds) Press button to enter calibration mode. (in 3 seconds) Press button to con rm present Temperature value. Press button twice to con rm selection.
  • Page 16: Communicating With A Pc

    10. COMMUNICATION WITH A PC Users can transfer data from the built-in memory on the instrument to a PC through SE376 software. The communication requires a USB or RS-232 serial connection. TESTLINK SE 376 SOFTWARE THE SE 376 PACKAGE CONTAINS: 80mm CD •...
  • Page 17 10. COMMUNICATION WITH A PC 6. Select “Launch the CP210x VCP Driver Installer” option, and then click Finish. (see Fig.2) (Fig.1) (Fig.2) 7. Click Install. 8. After nish installing USB-120 connector driver, it will install SE376 software automatically. it will copy SE376.exe ( executable le ) and help le to your hard disk ( default is c:\program les\SE376.
  • Page 18 10. COMMUNICATION WITH A PC 11. Choose “Hardware” option and click “Device Manager” button, click “+” symbol on the left side of Ports(COM &LPT), if the device “CP2101 USB to UART Bridge Controller(COMx)” is in the list, the connector is ready to work. MAIN MENU AND BUTTONS: Open - Retrieve les from the disk Save - Save the present data to the disk...
  • Page 19 10. COMMUNICATION WITH A PC Option - Setup dialog to set attributions of graph. (You can also double click on the graph.) Sampling Rate - Time interval between each data to be recorded. If you want to change it, move mouse cursor to the digit you want to change and click, then input the new sampling rate.
  • Page 20 10. COMMUNICATION WITH A PC TABULAR: The Tabular window can be used to record real time data in a table format. The maximum number of records that the Tabular window can keep will be depending on how many available memory and hard disk space of the connected computer.
  • Page 21 10. COMMUNICATION WITH A PC It will transfer the rst data set to graph and tabular on the right side window once you nish loading recorded data from the thermometer. You can click on at any data set to show its graph and tabular on the right side window.
  • Page 22 10. COMMUNICATION WITH A PC FREQUENTLY ASKED QUESTION: Q: I have connected thermometer to computer serial port and turned thermometer on, but it still shows "NO CONNECTION" ? A: It could be caused by that all serial port are occupied by other applications.
  • Page 23: Power Preparation

    11. POWER PREPARATION 11.1 BATTERY LOADING Remove the battery cover on the back and insert one 9V battery. 11.2 BATTERY REPLACEMENT When the battery voltage drops below the operating voltage, Low battery indicator will appear. Replace a 9 Volt battery immediately to ensure the unit functions properly 11.3 AC ADAPTER CONNECTION When the AC adapter is used, insert the plugs of the adapter into...
  • Page 24: Maintenance

    12. MAINTENANCE In order to ensure the accuracy of the thermometer for a longer period of time you should calibrate it once a year. Clean the device and the window of the display with a clean, lint-free, anti-static and dry cleaning cloth. Do not use cleaning agents that contain carbon or benzenes, alcohol or anything similar to clean the product since these substances damage the surface...
  • Page 25: Temperature Probe

    13. TEMPERATURE PROBE 13.1 PIERCING TYPE TEMPERATURE PROBE. 13.2 PIERCING TYPE TEMPERATURE PROBE SPECIFICATION: Sensor Type Platinum resistance thermometer sensor Pt 100(4 wires) Accuracy IEC751, class A ±0.15°C ±0.002t (t: measurement temperature) Measurement -100 to 400°C Range Temperature Sensor Approx. Ø5.0mm(Ø0.2”) Dimensions Temperature Sensor Approx.
  • Page 26 CUSTOMER SERVICE INQUIRIES Your emails are important to us so we strive to reply all hours. In exceptional cases, inquiries and emails within we may require more time to respond. Thank you for your understanding. For more information about our products and services, please send us an email: cs@perfectprime.com For B2B or project-based application, please send an email:...