Guidance and manufacturer's declaration - electromagnetic immunity
The PMW-10MD is intended for use in the electromagnetic environment specified below. The
customer or the user of the PMW-10MD should assure that it is used in such an environment.
Immunity test
Electrostatic
±6 kV contact
discharge (ESD)
IEC 61000-4-2
±8 kV air
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
±1 kV line(s) to
line(s)
IEC 61000-4-5
±2 kV line(s) to
earth
Voltage dips, short
< 5% U
interruptions and
(> 95% dip in U
voltage variations on
for 0.5 cycle
power supply input
lines
40% U
(60% dip in U
IEC 61000-4-11
for 5 cycles
70% U
(30% dip in U
for 25 cycles
< 5% U
(> 95% dip in U
for 5 sec
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601
Compliance level
test level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV differential
mode
±2 kV common
mode
< 5% U
T
T
)
(> 95% dip in U
T
for 0.5 cycle
40% U
T
T
)
(60% dip in U
T
for 5 cycles
70% U
T
T
)
(30% dip in U
T
for 25 cycles
< 5% U
T
T
)
(> 95% dip in U
T
for 5 sec
3 A/m
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
)
of a typical commercial or hospital
T
environment. If the user of the
PMW-10MD requires continued
operation during power mains
)
interruptions, it is recommended
T
that the PMW-10MD be powered
from an uninterruptible power
supply or a battery.
)
T
)
T
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.