Warming Up; Warming Blocks; Menu Function - K-Systems G95 Instruction Manual

Portable incubator
Table of Contents

Advertisement

5.8 Warming up

Warming up from 20.0 to 39. 0° C takes less than 15 minutes under normal conditions.
When a cold item is placed in the chamber it may need up to 30 minutes to stabilise,
depending on the thermal mass of the object and the initial temperature.
Always let the unit stabilise before use.

5.9 Warming blocks

It is necessary to use warming blocks. Please contact your local distributor or K-SYSTEMS for details
about the full range of warming blocks.
A selection of different tubes and dishes can be transported. Either tubes or dishes can be transported at
one time, not a combination of both.
WARNING
Warming blocks should always be pre-heated before usage.

6. Menu function

The user can access a number of advanced functions via the menu.
Action
Press and hold UP+DOWN key for 3
seconds to enter the the menu.
The first option "UNIT" will appear in the
display.
Press UP/DOWN key to navigate within the
Menu.
While holding the setpoint key in, press
either the arrow UP or arrow DOWN key to
change values within the options.
Press and hold UP+DOWN key for 3
seconds again to exit the menu.
Key
Page 12
Instruction manual - G95
Display
Kivex Biotec Ltd

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents