Microscope -> Incubation - Zeiss Axio Observer Series Operating Manual

Inverted microscope
Hide thumbs Also See for Axio Observer Series:
Table of Contents

Advertisement

ZEISS
Fig. 116
Microscope -> Incubation ->
Incubation page
Fig. 117
Pop-up window„H Insert P"
Fig. 118
Microscope -> Incubation -> Y
module page
130
TFT display touchscreen on the Axio Observer 7, 7 materials
431004-7244-001
OPERATION
5.11.8
Microscope -> Incubation
The Microscope -> Incubation page is only
available for Bio / Med stand types. It contains the
two tabs Incubation and Y module.
These tabs are used to control incubation
components and thermostats connected to the
microscope.
5.11.8.1
Incubation
The Incubation tab lists all fitted incubation
components. Each button shows the target
(bottom) and actual (top) temperatures for
temperature control.
Click on the appropriate button to call up the
corresponding pop-up window (e.g. H Insert P). In
this window you can switch the temperature
control on or off and set a target temperature
value.
5.11.8.2
Y module
The Y module tab lists the thermostats connected
to the microscope. Each button shows the target
(bottom) and actual (top) temperatures for
temperature control.
In the same way as for incubation components,
thermostats can be switched on or off and target
temperatures can be set from the pop-up
thermostat window.
Axio Observer