Lubrication Of Chucking System - W&H assistina 301 plus Instructions For Use Manual

Hide thumbs Also See for assistina 301 plus:
Table of Contents

Advertisement

Operation
FG contra-angle handpieces
> Remove the rotary instrument.
> Use the adaptor 02693000
> Wear protective clothing, saftey glasses, face mask and gloves.
24
With its spray exit nozzle
directed downwards, push the
adaptor onto the maintenance
attachment.
Initially, keep the cover open.
Push the transmission
instrument onto the adaptor
from underneath. Support the
transmission instrument at
the edge of the cover.

Lubrication of chucking system

1 x
2 sec.
Press the start button once for
about 2 seconds.
Press the transmission
instrument against the
adaptor for about
10 seconds.
Close the cover. Allow the
remainder of the maintenance
cycle to complete.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents