Keysight Technologies 346A Installation Note
Keysight Technologies 346A Installation Note

Keysight Technologies 346A Installation Note

Noise sources instructions for setting bias current

Advertisement

Quick Links

Keysight Technologies
Noise Sources:
346A, 346B, N4000A, and
N4001A (All Serial Numbers)
Instructions for Setting Bias
Current
Installation
Note

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 346A and is the answer not in the manual?

Questions and answers

Summary of Contents for Keysight Technologies 346A

  • Page 1 Keysight Technologies Noise Sources: 346A, 346B, N4000A, and N4001A (All Serial Numbers) Instructions for Setting Bias Current Installation Note...
  • Page 2 FAR 27.401 or DFAR 227.7103-5 (c), adhered to, could result in personal injury or as applicable in any technical data. death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met. Keysight 346A, 346B, N4000A, and N4001A Installation Note...
  • Page 3: Product Category

    The information contained in this document is subject to change without notice. Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 4: Waste Electrical And Electronic Equipment (Weee) Directive

    To contact Keysight for sales and technical support, refer to the support links on the following Keysight websites: – www.keysight.com/find/mta (product-specific information and support, software and documentation updates) – www.keysight.com/find/assist (worldwide contact information for repair and service) Keysight 346A, 346B, N4000A, and N4001A Installation Note...
  • Page 5: Table Of Contents

    346A/B Replacement Bulkhead Assembly Installation Instructions ....13 346A/B Replacement Printed Circuit Board Assembly Installation Instructions ..14 N4000A/ N4001A Replacement Noise Cartridge Installation Instructions .
  • Page 6 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 346A, 346B, N4000A, and N4001A Installation Note...
  • Page 7: List Of Figures

    ..........16 Keysight 346A, 346B, N4000A, and N4001A Installation Note...
  • Page 8 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 346A, 346B, N4000A, and N4001A Installation Note...
  • Page 9: List Of Tables

    ..........16 Keysight 346A, 346B, N4000A, and N4001A Installation Note...
  • Page 10 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 346A, 346B, N4000A, and N4001A Installation Note...
  • Page 11: Introduction

    The noise source diode inside the noise source cartridge changed at serial prefix MY/SG/US5323 for models 346A and N4000A including options 001, 002, and 004. Noise sources with serial prefixes manufactured before MY/SG/US5323 will now have a noise cartridge replacement kit that will include both a new noise cartridge and new attenuator.
  • Page 12: 346A/B Replacement Noise Cartridge Installation Instructions

    ENR specification. For the 346A and 346B, bias levels are determined by installing the appropriate resistor value for R7 on the printed circuit board (Figure 1).
  • Page 13: 346A/B Replacement Bulkhead Assembly Installation Instructions

    12 For the 346B, if the ENR is not within specification, refer to the supplied 00346-90024 Installation Note for instructions on how to tune the attenuator to meet the ENR specification at frequencies above 12 GHz. The 346A should not require polyiron tuning. 13 Retest ENR after tuning the attenuator.
  • Page 14: 346A/B Replacement Printed Circuit Board Assembly Installation Instructions

    346A/B Replacement Printed Circuit Board Assembly Installation Instructions 1 Disassemble the Noise Source by removing the two flat head screws near the BNC connector, and slide out the printed circuit board assembly. 2 R7 is not loaded in the replacement printed circuit board assembly. Install the same value of R7 as in the printed circuit board being replaced.
  • Page 15: N4000A/ N4001A Replacement Bulkhead Assembly Installation Instructions

    7 If the ENR is not meeting specification, use the field calibration software to change the bias setting. Refer to Table 2 for bias current values. Decreasing the bias current increases the ENR response. 8 Retest ENR. Keysight 346A, 346B, N4000A, and N4001A Installation Note...
  • Page 16: Printed Circuit Board Assembly Graphic And Resistor List

    Parts list Current (mA) R7 ohms R7 part number 0698-3442 0698-6377 0757-0405 0698-7215 0757-0401 0698-7209 51.1 0757-0394 28.7 0698-7199 0757-0346 0698-7217 [a] In addition, replace R6 with 0698-7216 (147 ohms) Figure 1 00346-60183 Keysight 346A, 346B, N4000A, and N4001A Installation Note...
  • Page 17 This information is subject to change without notice. Always refer to the Keysight website for the latest revision. © Keysight Technologies 2001 - 2019 Edition 3, December 16, 2019 Printed in Malaysia *00346-90137* 00346-90137 www.keysight.com...

This manual is also suitable for: