Siemens SITRANS FUP1010 Quick Start Manual page 26

Ip67 portable ultrasonic flowmeter
Hide thumbs Also See for SITRANS FUP1010:
Table of Contents

Advertisement

Commissioning
3.3 Setting the Parameters
5. The drop down menu lists the following sensor selections:
6. For this example, select the sensor model that appears on the sensor label then press
7. To select Sensor Size, press <Right Arrow>. Scroll to select the sensor size that matches
8. At [Sensor Mount Mode], press the <Right Arrow>. Scroll to select [Reflect] or [Direct]
9. IMPORTANT: Record Spacing Method and Number Index. This data will be used to
10. Sensors can now be mounted. Refer to Sensor Installation (Page 25) mounting
11. After sensors are mounted scroll to [Install Complete] and select [Install].
24
– 1011 Universal
– 1011HP-T1 - Usable -40 to 120°C, recommended for Ø Temperature <40°C;
Standard.
– 1011HP-T2 - Usable -40 to 120°C, recommended for Ø Temperature >40°C - <80°C;
Named as high temperature.
– 1011HP-T3 - Usable -40 to 120°C, recommended for Ø Temperature >80°C <120°C;
special request.
– 991 Universal
Note
The meter will automatically recommend a sensor depending on the application data
that has been entered.
<ENTER>.
Select based on type.
Select based on size.
After Sensor is mounted select [Install].
the size indicated on the sensor label. Press <ENTER>.
mount and then press <ENTER>.
mount the sensors.
procedures and select the mounting mode desired.
FUP1010 IP67 Portable Quick Start
Operating Instructions, 01/2013, CQO:QSG003 Revision 05

Advertisement

Table of Contents
loading

Table of Contents