Download Print this page

Specifications; Battery Replacement; Operation; Meter Setup - Hanna Instruments Checktemp 1 Instruction Manual

Advertisement

Thank you!
Thank you for choosing a Hanna Instruments product. Please read this instruction manual
carefully before using the instrument. For additional information, visit www.hannainst.com or
e-mail us at tech@hannainst.com.

Specifications

Range
-50.0 to 150.0 ºC / -58.0 to 302 ºF
0.1 ºC (-50.0 to 150.0 ºC)
Resolution
0.1 ºF (-58.0 to 199.9 ºF) / 1 ºF (above 200 ºF)
±0.2 ºC (-30.0 to 120.0 ºC)
±0.3 ºC (outside: -50.0 to -30.0 ºC and 120.0 to 150.0 ºC)
Accuracy
±0.5 ºF (-22.0 to 199.9 ºF)
±1 ºF (outside: -58.0 to -22.0 ºF and 200 to 302 ºF)
Probe
stainless steel probe with 1 m (3.3')
Environment
-30 to 50 ºC (-22 to 122 ºF)
Battery Type
1.5V AAA (3 pcs.)
Battery Life
approx. 2 years of use
Auto Off
8 min (default), 60 min. or OFF
Dimensions
107 x 59 x 17 mm (4.2 x 5.3 x 0.7")
Weight
130 g (4.6 oz.)
EN 13485:2001 - Suitability: Storage;
Certification
Environment: E; Accuracy Class: 0.5; IP65

Operation

Turn Checktemp
ON with ON/OFF button: all LCD segments are displayed for 1 second, and
®
CalCheck
starts automatically to verify that Checktemp
®
message of "CAL" and "-0-" means that Checktemp
used. If CalCheck
displays the "Err" message while measuring temperature, remember that:
®
• The temperature sensor is located at the tip of the stainless steel probe.
• It is necessary to wait for a few seconds for the display to stabilize.

Meter Setup

Enter Setup to change the temperature unit (ºC or ºF) or to adjust the Auto Shut Off time. To
enter the Setup MODE, press the SET button located inside the battery compartment once for
temperature unit or twice for Auto Shut Off.
• Use the ON/OFF button to switch between "ºC" or "ºF". Press SET once to save.
• Use the ON/OFF button to cycle through the following Auto Shut Off options: 8 minutes
("d08", default value), 60 minutes ("d60"), OFF ("d--"). Press the SET button once to save.
is still accurate to its specifications. A
®
is within specifications and ready to be
®

Battery Replacement

Replace the battery when the battery symbol is blinking or when Checktemp
on. Open the compartment and replace the three AAA batteries, while paying attention to their
polarity. Close the battery cover.
Note: Batteries should only be replaced in a safe area using the battery type specified in this
instruction manual. Old batteries should be disposed in accordance with local regulations.
REAR
Battery
cover
Meter stand
Warranty
Checktemp
is guaranteed for a period of one year after date of purchase against defects in work-
®
manship and materials when used for their intended purpose and maintained according to
instructions. This warranty is limited to repair or replacement free of charge. Damage due to accidents, mis-
use, tampering or lack of prescribed maintenance is not covered. If service is required, contact your local Hanna
Instruments Office. If under warranty, report the model number, date of purchase, serial number and the nature of
problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument
is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Technical
Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is
properly packed for complete protection.
Recommendations for Users
Before using Hanna Instruments products, make sure that they are entirely suitable for your specific application and
for the environment in which they are used. Any variation introduced by the user to the supplied equipment may
degrade the meter's performance. Batteries should only be replaced in a safe area using the battery type specified
in this instruction manual. To avoid damages or burns, do not put the meter in microwave oven. For yours and the
meter's safety do not use or store the meter in hazardous environments.
1 does not turn
®
Silicone Probe Cable
Liquid Crystal Display
Maximum Immersion Level
Temperature Probe
4 cm
Recommended Immersion Level
1.5"
IST98509 06/18
ON/OFF
button
SET
button

Advertisement