Trouble Shooting - M&C CG Series Instruction Manual

Gas converter
Table of Contents

Advertisement

20

TROUBLE SHOOTING

The following table is intended to show the possible sources of errors and their rectification
Problem/Display
LED's do not light up
Valves do not switch over
Temperature controller out of
order
Converter does not heat up
LED K1 off
Valves do not switch
LED's do not light up
Valves do not switch
LED's light up
No sample gas flow
No conversion
www.mc-techgroup.com
Possible Cause
No mains power
Sub-D-plug not inserted in
socket X2
F1/2 fuses defective
Heater defective
Temperature controller
defective
Solid-state relay defective
Sensor short circuit
No mains supply (see above)
Sub-D-plug not inserted into
socket X2 (see above)
Internal circuit;
No solder link 1-6 in sub-D-
plug
External circuit;
Error in external controller
Valve defective
Valves defective (see above)
Gas sample lines
blocked or leaking
Cartridge used up
Cartridge does not heat up
(see above)
No sample gas flow (see
above)
CG-2 | 1.04.00
Check/Rectification
Mains 'ON'
OK.
Check that mains cable fits properly
(X1);
OK.
Check whether sub-D-plug is present
and is properly plugged in;
OK.
Check whether bridge in sub-D-plug ex-
ists (s. Fig. 6);
OK.
Check fuses and replace if necessary;
Measure voltage at terminals X4/2 and
3;
OK.
Replace heater;
Not OK.
Measure voltage at terminal X4/6 and
7;
Voltage < 8 V DC?
Check controller according to operating
instructions;
Voltage > 8 V DC?
Replace solid-state relay;
Change sensor;
see above
see above
Check sub-D-plug and if necessary sol-
der link;
Check external controller;
Check that valves function
See above
Check gas sample lines;
Check for gas-tightness
Change cartridge (see chapter 17.1)
See above
See above